Praz, Fabien; Braun, Daniel; Unterhuber, Matthias; Spirito, Alessandro; Orban, Mathias; Brugger, Nicolas; Brinkmann, Isabel; Spring, Karin; Moschovitis, Aris; Nabauer, Michael; Blazek, Stephan; Pilgrim, Thomas; Thiele, Holger; Lurz, Philipp; Hausleiter, Jörg; Windecker, Stephan (2019). Edge-to-Edge Mitral Valve Repair With Extended Clip Arms: Early Experience From a Multicenter Observational Study. JACC. Cardiovascular Interventions, 12(14), pp. 1356-1365. Elsevier 10.1016/j.jcin.2019.03.023

[img] Text
Edge-to-Edge Mitral Valve Repair With Extended Clip Arms.pdf - Published Version
Restricted to registered users only
Available under License Publisher holds Copyright.

Download (1MB)

OBJECTIVES

The aim of this study was to investigate the technical success and efficacy of mitral valve edge-to-edge repair using extended clip arms.

BACKGROUND

A new iteration of the MitraClip system, the MitraClip XTR, was introduced in 2018 with the aim of addressing technical limitations observed with previous versions.

METHODS

Patients having received at least 1 new implant for the treatment of symptomatic mitral regurgitation (MR) were eligible for this study.

RESULTS

Among the 107 patients (mean age 76 ± 9 years, 69% men) included in this study, the etiology of MR was balanced, with one-half (n = 53 [50%]) classified as secondary and the remaining 54 patients having either primary (n = 40 [37%]) or mixed (n = 14 [13%]) disease. The mean number of devices implanted was 1.5 ± 0.6. Multiple device implantation was required in 46 patients (43%). Single-leaflet device attachment occurred in 4 patients and leaflet injury in 2 additional patients, requiring surgical conversion in 4 patients. Among the 102 patients discharged alive without mitral valve surgery, 95 (93%) had MR ≤2+ and 79 (77%) had MR ≤1+. The mean transmitral gradient increased from 1.9 ± 1.0 mm Hg at baseline to 3.5 ± 1.8 mm Hg at discharge (p < 0.001).

CONCLUSIONS

Technical success with the new mitral valve repair system with extended clip arm was achieved in 93% of the patients. MR ≤2+ was obtained in 95 patients (93%) and MR ≤1+ in 79 (77%). The main reasons for procedural failure were acute single-leaflet device attachment associated with leaflet damage or isolated leaflet injury and often required surgical correction.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Cardiovascular Disorders (DHGE) > Clinic of Cardiology

UniBE Contributor:

Praz, Fabien Daniel, Pilgrim, Thomas, Windecker, Stephan

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1876-7605

Publisher:

Elsevier

Language:

English

Submitter:

Nadia Biscozzo

Date Deposited:

11 Feb 2020 08:35

Last Modified:

02 Mar 2023 23:33

Publisher DOI:

10.1016/j.jcin.2019.03.023

PubMed ID:

31129091

Uncontrolled Keywords:

MitraClip mitral regurgitation mitral valve transcatheter mitral valve repair