Praz, Fabien; Braun, Daniel; Unterhuber, Matthias; Spirito, Alessandro; Orban, Mathias; Brugger, Nicolas; Brinkmann, Isabel; Spring, Karin; Moschovitis, Aris; Nabauer, Michael; Blazek, Stephan; Pilgrim, Thomas; Thiele, Holger; Lurz, Philipp; Hausleiter, Jörg; Windecker, Stephan (2019). Edge-to-Edge Mitral Valve Repair With Extended Clip Arms: Early Experience From a Multicenter Observational Study. JACC. Cardiovascular Interventions, 12(14), pp. 1356-1365. Elsevier 10.1016/j.jcin.2019.03.023
Text
Edge-to-Edge Mitral Valve Repair With Extended Clip Arms.pdf - Published Version Restricted to registered users only Available under License Publisher holds Copyright. Download (1MB) |
OBJECTIVES
The aim of this study was to investigate the technical success and efficacy of mitral valve edge-to-edge repair using extended clip arms.
BACKGROUND
A new iteration of the MitraClip system, the MitraClip XTR, was introduced in 2018 with the aim of addressing technical limitations observed with previous versions.
METHODS
Patients having received at least 1 new implant for the treatment of symptomatic mitral regurgitation (MR) were eligible for this study.
RESULTS
Among the 107 patients (mean age 76 ± 9 years, 69% men) included in this study, the etiology of MR was balanced, with one-half (n = 53 [50%]) classified as secondary and the remaining 54 patients having either primary (n = 40 [37%]) or mixed (n = 14 [13%]) disease. The mean number of devices implanted was 1.5 ± 0.6. Multiple device implantation was required in 46 patients (43%). Single-leaflet device attachment occurred in 4 patients and leaflet injury in 2 additional patients, requiring surgical conversion in 4 patients. Among the 102 patients discharged alive without mitral valve surgery, 95 (93%) had MR ≤2+ and 79 (77%) had MR ≤1+. The mean transmitral gradient increased from 1.9 ± 1.0 mm Hg at baseline to 3.5 ± 1.8 mm Hg at discharge (p < 0.001).
CONCLUSIONS
Technical success with the new mitral valve repair system with extended clip arm was achieved in 93% of the patients. MR ≤2+ was obtained in 95 patients (93%) and MR ≤1+ in 79 (77%). The main reasons for procedural failure were acute single-leaflet device attachment associated with leaflet damage or isolated leaflet injury and often required surgical correction.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > Department of Cardiovascular Disorders (DHGE) > Clinic of Cardiology |
UniBE Contributor: |
Praz, Fabien Daniel, Pilgrim, Thomas, Windecker, Stephan |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
1876-7605 |
Publisher: |
Elsevier |
Language: |
English |
Submitter: |
Nadia Biscozzo |
Date Deposited: |
11 Feb 2020 08:35 |
Last Modified: |
02 Mar 2023 23:33 |
Publisher DOI: |
10.1016/j.jcin.2019.03.023 |
PubMed ID: |
31129091 |
Uncontrolled Keywords: |
MitraClip mitral regurgitation mitral valve transcatheter mitral valve repair |