Raillard, Mathieu; Levionnois, Olivier; MacFarlane, Paul (2019). Do the Manual or Computer-Controlled Flowmeters Generate Similar Isoflurane Concentrations in Tafonius? Frontiers in veterinary science, 6 Frontiers Media 10.3389/fvets.2019.00160
|
Text
fvets-06-00160.pdf - Published Version Available under License Creative Commons: Attribution (CC-BY). Download (821kB) | Preview |
Introduction: Tafonius is an anesthesia machine with computer-controlled monitor and ventilator. We compared the isoflurane fluctuations in the circuit with manual (MF) or computer-driven (CF) flowmeters, investigated the origin of the differences and assessed whether isoflurane concentration time course followed a one-compartment model.
Material and Methods: A calibrated TEC-3 isoflurane vaporizer was used. Gas composition and flows were measured using a multiparametric monitor and a digital flowmeter. Measurements included: (1) Effects of various FiO2 with MF/CF on the isoflurane fraction changes in the breathing system during mechanical ventilation of a lung model; wash-in kinetic was fitted to a compartmental model; (2) Gas outflow at the common gas outlet (CGO) with MF/CF at different FiO2; (3) Isoflurane output of the vaporizer at various dial settings with MF/CF set at different flows without and with reduction of the CGO diameter.
Results: (1) The 3% targeted isoflurane concentration was not reached; additional time was required to reach specific concentrations with CF (lowest FiO2, longer time). The exponential course fitted a two-compartment model; (2) Set and measured flows were identical with MF. With CF at 0.21 FiO2, flow was intermittently 7.6 L min−1 or zero (mean total: 38% of the set flow); with CF at 1.00 FiO2, flow was 10.6 L min−1 or zero (mean: 4–5.3 L min−1); with 0.21 < FiO2 < 1.00, combined flow was intermittent (maximum output: 15.6 L min−1); (3) With MF, isoflurane output was matching dial setting at 5 L min−1 but was lower at higher flows; with CF generating intermittent flows, isoflurane output was fluctuating. With the 4 mm diameter CGO, isoflurane concentration was close to dial setting with both MF and CF. With a 14 G CGO, isoflurane concentration was lower than dial setting with MF, higher with CF.
Conclusions and Clinical Relevance: Using MF or CF led to different isoflurane fraction time course in Tafonius. Flows were lower than set with CF; the TEC-3 did not compensate for high/intermittent flows and pressures; the CGO diameter influenced isoflurane output.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
05 Veterinary Medicine > Department of Clinical Veterinary Medicine (DKV) > DKV - Anaesthesiology 05 Veterinary Medicine > Department of Clinical Veterinary Medicine (DKV) |
UniBE Contributor: |
Raillard, Mathieu Clément, Levionnois, Olivier |
Subjects: |
600 Technology > 630 Agriculture 600 Technology > 610 Medicine & health |
ISSN: |
2297-1769 |
Publisher: |
Frontiers Media |
Language: |
English |
Submitter: |
Claudia Spadavecchia |
Date Deposited: |
04 Feb 2020 11:03 |
Last Modified: |
17 Apr 2024 16:13 |
Publisher DOI: |
10.3389/fvets.2019.00160 |
BORIS DOI: |
10.7892/boris.139351 |
URI: |
https://boris.unibe.ch/id/eprint/139351 |