Using a Bone Conduction Hearing Device as a Tactile Aid.

Kompis, Martin; Langmair, Manfred; Mantokoudis, Georgios; Weder, Stefan; Gawliczek, Tom; Caversaccio, Marco (2023). Using a Bone Conduction Hearing Device as a Tactile Aid. Audiology research, 13(3), pp. 459-465. MDPI 10.3390/audiolres13030040

[img]
Preview
Text
audiolres-13-00040.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (1MB) | Preview

BACKGROUND

With the advent of cochlear implants, tactile aids for the profoundly deaf became obsolete decades ago. Nevertheless, they might still be useful in rare cases. We report the case of a 25-year-old woman with Bosley-Salih-Alorainy Syndrome and bilateral cochlear aplasia.

METHODS

After it was determined that cochlear or brainstem implants were not an option and tactile aids were not available anymore, a bone conduction device (BCD) on a softband was tried as a tactile aid. The usual retroauricular position and a second position close to the wrist, preferred by the patient, were compared. Sound detection thresholds were measured with and without the aid. Additionally, three bilaterally deaf adult cochlear implant users were tested under the same conditions.

RESULTS

At 250-1000 Hz, sounds were perceived as vibrations above approximately 45-60 dB with the device at the wrist. Thresholds were approximately 10 dB poorer when placed retroauricularly. Differentiation between different sounds seemed difficult. Nevertheless, the patient uses the device and can perceive loud sounds.

CONCLUSIONS

Cases where the use of tactile aids may make sense are probably very rare. The use of BCD, placed, e.g., at the wrist, may be useful, but sound perception is limited to low frequencies and relatively loud levels.

Item Type:

Journal Article (Further Contribution)

Division/Institute:

04 Faculty of Medicine > Department of Head Organs and Neurology (DKNS) > Clinic of Ear, Nose and Throat Disorders (ENT)

UniBE Contributor:

Kompis, Martin, Mantokoudis, Georgios, Weder, Stefan Andreas, Gawliczek, Tom, Caversaccio, Marco

Subjects:

600 Technology > 610 Medicine & health

ISSN:

2039-4330

Publisher:

MDPI

Language:

English

Submitter:

Pubmed Import

Date Deposited:

28 Jun 2023 09:46

Last Modified:

28 Jun 2023 10:24

Publisher DOI:

10.3390/audiolres13030040

PubMed ID:

37366686

Uncontrolled Keywords:

bone conduction cochlear aplasia sound field sound processor tactile aids vibratory sensation