Fatigue life of 3D-printed porous titanium dental implants predicted by validated finite element simulations.

Vautrin, Antoine; Aw, Jensen; Attenborough, Ed; Varga, Peter (2023). Fatigue life of 3D-printed porous titanium dental implants predicted by validated finite element simulations. Frontiers in Bioengineering and Biotechnology, 11(1240125), p. 1240125. Frontiers Media 10.3389/fbioe.2023.1240125

[img]
Preview
Text
fbioe-11-1240125.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (2MB) | Preview

Introduction: Porous dental implants represent a promising strategy to reduce failure rate by favoring osseointegration or delivering drugs locally. Incorporating porous features weakens the mechanical capacity of an implant, but sufficient fatigue strength must be ensured as regulated in the ISO 14801 standard. Experimental fatigue testing is a costly and time-intensive part of the implant development process that could be accelerated with validated computer simulations. This study aimed at developing, calibrating, and validating a numerical workflow to predict fatigue strength on six porous configurations of a simplified implant geometry. Methods: Mechanical testing was performed on 3D-printed titanium samples to establish a direct link between endurance limit (i.e., infinite fatigue life) and monotonic load to failure, and a finite element model was developed and calibrated to predict the latter. The tool was then validated by predicting the fatigue life of a given porous configuration. Results: The normalized endurance limit (10% of the ultimate load) was the same for all six porous designs, indicating that monotonic testing was a good surrogate for endurance limit. The geometry input of the simulations influenced greatly their accuracy. Utilizing the as-designed model resulted in the highest prediction error (23%) and low correlation between the estimated and experimental loads to failure (R2 = 0.65). The prediction error was smaller when utilizing specimen geometry based on micro computed tomography scans (14%) or design models adjusted to match the printed porosity (8%). Discussion: The validated numerical workflow presented in this study could therefore be used to quantitatively predict the fatigue life of a porous implant, provided that the effect of manufacturing on implant geometry is accounted for.

Item Type:

Journal Article (Original Article)

Graduate School:

Graduate School for Cellular and Biomedical Sciences (GCB)

ISSN:

2296-4185

Publisher:

Frontiers Media

Language:

English

Submitter:

Pubmed Import

Date Deposited:

28 Aug 2023 16:11

Last Modified:

24 Sep 2023 02:34

Publisher DOI:

10.3389/fbioe.2023.1240125

PubMed ID:

37636001

Uncontrolled Keywords:

additive manufacturing dental implant fatigue finite element analysis micro-CT porous titanium

BORIS DOI:

10.48350/185807

URI:

https://boris.unibe.ch/id/eprint/185807

Actions (login required)

Edit item Edit item
Provide Feedback