Alkojak Almansi, Abdulrahman; Sugarova, Sima; Alsanosi, Abdulrahman; Almuhawas, Fida; Hofmeyr, Louis; Wagner, Franca; Kedves, Emerencia; Sriperumbudur, Kiran; Dhanasingh, Anandhan; Kedves, Andras (2024). A novel radiological software prototype for automatically detecting the inner ear and classifying normal from malformed anatomy. Computers in biology and medicine, 171 Elsevier 10.1016/j.compbiomed.2024.108168

[img]
Preview
Text
1-s2.0-S001048252400252X-main.pdf - Published Version
Available under License Creative Commons: Attribution-Noncommercial-No Derivative Works (CC-BY-NC-ND).

Download (3MB) | Preview

BACKGROUND

To develop an effective radiological software prototype that could read Digital Imaging and Communications in Medicine (DICOM) files, crop the inner ear automatically based on head computed tomography (CT), and classify normal and inner ear malformation (IEM).

METHODS

A retrospective analysis was conducted on 2053 patients from 3 hospitals. We extracted 1200 inner ear CTs for importing, cropping, and training, testing, and validating an artificial intelligence (AI) model. Automated cropping algorithms based on CTs were developed to precisely isolate the inner ear volume. Additionally, a simple graphical user interface (GUI) was implemented for user interaction. Using cropped CTs as input, a deep learning convolutional neural network (DL CNN) with 5-fold cross-validation was used to classify inner ear anatomy as normal or abnormal. Five specific IEM types (cochlear hypoplasia, ossification, incomplete partition types I and III, and common cavity) were included, with data equally distributed between classes. Both the cropping tool and the AI model were extensively validated.

RESULTS

The newly developed DICOM viewer/software successfully achieved its objectives: reading CT files, automatically cropping inner ear volumes, and classifying them as normal or malformed. The cropping tool demonstrated an average accuracy of 92.25%. The DL CNN model achieved an area under the curve (AUC) of 0.86 (95% confidence interval: 0.81-0.91). Performance metrics for the AI model were: accuracy (0.812), precision (0.791), recall (0.8), and F1-score (0.766).

CONCLUSION

This study successfully developed and validated a fully automated workflow for classifying normal versus abnormal inner ear anatomy using a combination of advanced image processing and deep learning techniques. The tool exhibited good diagnostic accuracy, suggesting its potential application in risk stratification. However, it is crucial to emphasize the need for supervision by qualified medical professionals when utilizing this tool for clinical decision-making.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Institute of Diagnostic and Interventional Neuroradiology

UniBE Contributor:

Wagner, Franca

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1879-0534

Publisher:

Elsevier

Language:

English

Submitter:

Martin Zbinden

Date Deposited:

29 Apr 2024 14:49

Last Modified:

29 Apr 2024 14:58

Publisher DOI:

10.1016/j.compbiomed.2024.108168

PubMed ID:

38432006

Uncontrolled Keywords:

Artificial intelligence Deep learning Digital health Radiology

BORIS DOI:

10.48350/196336

URI:

https://boris.unibe.ch/id/eprint/196336

Actions (login required)

Edit item Edit item
Provide Feedback