The relation between aortic morphology and transcatheter aortic heart valve thrombosis: Particle tracing and platelet activation in larger aortic roots with and without neo-sinus.

Bornemann, Karoline-Marie; Jahren, Silje Ekroll; Obrist, Dominik (2024). The relation between aortic morphology and transcatheter aortic heart valve thrombosis: Particle tracing and platelet activation in larger aortic roots with and without neo-sinus. Computers in biology and medicine, 179, p. 108828. Elsevier 10.1016/j.compbiomed.2024.108828

[img]
Preview
Text
1-s2.0-S0010482524009132-main.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (1MB) | Preview

Transcatheter aortic heart valve thrombosis (THVT) affects long-term valve durability, transvalvular pressure gradient and leaflet mobility. In this study, we conduct high-fidelity fluid-structure interaction simulations to perform Lagrangian particle tracing in a generic model with larger aortic diameters (THVT model) with and without neo-sinus which is compared to a model of unaffected TAVI patients (control model). Platelet activation indices are computed for each particle to assess the risk of thrombus formation induced by high shear stresses followed by flow stagnation. Particle tracing indicates that fewer particles contribute to sinus washout of the THVT model with and without neo-sinus compared to the control model (-34.9%/-34.1%). Stagnating particles in the native sinus of the THVT model show higher platelet activation indices than for the control model (+39.6% without neo-sinus, +45.3% with neo-sinus). Highest activation indices are present for particles stagnating in the neo-sinus of the larger aorta representing THVT patients (+80.2% compared to control). This fluid-structure interaction (FSI) study suggests that larger aortas lead to less efficient sinus washout in combination with higher risk of platelet activation among stagnating particles, especially within the neo-sinus. This could explain (a) a higher occurrence of thrombus formation in transcatheter valves compared to surgical valves without neo-sinus and (b) the neo-sinus as the prevalent region for thrombi in TAV. Pre-procedural identification of larger aortic roots could contribute to better risk assessment of patients and improved selection of a patient-specific anti-coagulation therapy.

Item Type:

Journal Article (Original Article)

Division/Institute:

10 Strategic Research Centers > ARTORG Center for Biomedical Engineering Research > ARTORG Center - Cardiovascular Engineering (CVE)
10 Strategic Research Centers > ARTORG Center for Biomedical Engineering Research

Graduate School:

Graduate School for Cellular and Biomedical Sciences (GCB)

UniBE Contributor:

Bornemann, Karoline-Marie, Jahren, Silje Ekroll, Obrist, Dominik

Subjects:

500 Science > 570 Life sciences; biology
600 Technology > 610 Medicine & health
600 Technology > 620 Engineering

ISSN:

1879-0534

Publisher:

Elsevier

Language:

English

Submitter:

Pubmed Import

Date Deposited:

17 Jul 2024 11:12

Last Modified:

05 Aug 2024 12:03

Publisher DOI:

10.1016/j.compbiomed.2024.108828

PubMed ID:

38996554

Uncontrolled Keywords:

Blood damage index Computational modeling Direct Numerical Simulation Flow stagnation Fluid–structure interaction Shear stress Sinus washout

BORIS DOI:

10.48350/198997

URI:

https://boris.unibe.ch/id/eprint/198997

Actions (login required)

Edit item Edit item
Provide Feedback