van Ingen, Jakko; Kohl, Thomas A; Kranzer, Katharina; Hasse, Barbara; Keller, Peter M; Katarzyna Szafrańska, Anna; Hillemann, Doris; Chand, Meera; Schreiber, Peter Werner; Sommerstein, Rami; Berger, Christoph; Genoni, Michele; Rüegg, Christian; Troillet, Nicolas; Widmer, Andreas F; Becker, Sören L; Herrmann, Mathias; Eckmanns, Tim; Haller, Sebastian; Höller, Christiane; ... (2017). Global outbreak of severe Mycobacterium chimaera disease after cardiac surgery: a molecular epidemiological study. The lancet. Infectious diseases, 17(10), pp. 1033-1041. Elsevier 10.1016/S1473-3099(17)30324-9

[img] Text
1-s2.0-S1473309917303249-main.pdf - Published Version
Restricted to registered users only
Available under License Publisher holds Copyright.

Download (2MB)

BACKGROUND

Since 2013, over 100 cases of Mycobacterium chimaera prosthetic valve endocarditis and disseminated disease were notified in Europe and the USA, linked to contaminated heater-cooler units (HCUs) used during cardiac surgery. We did a molecular epidemiological investigation to establish the source of these patients' disease.

METHODS

We included 24 M chimaera isolates from 21 cardiac surgery-related patients in Switzerland, Germany, the Netherlands, and the UK, 218 M chimaera isolates from various types of HCUs in hospitals, from LivaNova (formerly Sorin; London, UK) and Maquet (Rastatt, Germany) brand HCU production sites, and unrelated environmental sources and patients, as well as eight Mycobacterium intracellulare isolates. Isolates were analysed by next-generation whole-genome sequencing using Illumina and Pacific Biosciences technologies, and compared with published M chimaera genomes.

FINDINGS

Phylogenetic analysis based on whole-genome sequencing of 250 isolates revealed two major M chimaera groups. Cardiac surgery-related patient isolates were all classified into group 1, in which all, except one, formed a distinct subgroup. This subgroup also comprised isolates from 11 cardiac surgery-related patients reported from the USA, most isolates from LivaNova HCUs, and one from their production site. Isolates from other HCUs and unrelated patients were more widely distributed in the phylogenetic tree.

INTERPRETATION

HCU contamination with M chimaera at the LivaNova factory seems a likely source for cardiothoracic surgery-related severe M chimaera infections diagnosed in Switzerland, Germany, the Netherlands, the UK, the USA, and Australia. Protective measures and heightened clinician awareness are essential to guarantee patient safety.

FUNDING

Partly funded by the EU Horizon 2020 programme, its FP7 programme, the German Center for Infection Research (DZIF), the Swiss National Science Foundation, the Swiss Federal Office of Public Health, and National Institute of Health Research Oxford Health Protection Research Units on Healthcare Associated Infection and Antimicrobial Resistance.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Haematology, Oncology, Infectious Diseases, Laboratory Medicine and Hospital Pharmacy (DOLS) > Clinic of Infectiology

UniBE Contributor:

Sommerstein, Rami

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1474-4457

Publisher:

Elsevier

Language:

English

Submitter:

Annelies Luginbühl

Date Deposited:

10 Nov 2017 15:42

Last Modified:

05 Dec 2022 15:06

Publisher DOI:

10.1016/S1473-3099(17)30324-9

PubMed ID:

28711585

BORIS DOI:

10.7892/boris.102459

URI:

https://boris.unibe.ch/id/eprint/102459

Actions (login required)

Edit item Edit item
Provide Feedback