Rapid immunoassays for diagnosis of heparin-induced thrombocytopenia: Comparison of diagnostic accuracy, reproducibility, and costs in clinical practice.

Bankova, Andriyana; Andres, Yvonne; Horn, Michael Peter; Alberio, Lorenzo; Nagler, Michael (2017). Rapid immunoassays for diagnosis of heparin-induced thrombocytopenia: Comparison of diagnostic accuracy, reproducibility, and costs in clinical practice. PLoS ONE, 12(6), e0178289. Public Library of Science 10.1371/journal.pone.0178289

[img]
Preview
Text
MN_Rapid immunoassays for diagnosis of heparin-induced thrombocytopenia.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (2MB) | Preview

BACKGROUND

Immunoassays are crucial in the work-up of patients with suspected heparin-induced thrombocytopenia (HIT) and rapid tests have been recently developed. However, comparative data on diagnostic accuracy, reproducibility, and analytical costs of different immunoassays in clinical practice are limited.

METHODS

Samples of 179 consecutive patients evaluated for suspected HIT in clinical practice using a polyspecific enzyme-linked immunoabsorbent assay (GTI diagnostics; ELISA) and a rapid particle gel immunoassay (PaGIA), were additionally analysed with a IgG-specific chemiluminescent immunoassay (AcuStar HIT-IgG). Presence of HIT was defined as a positive functional heparin-induced platelet aggregation test. Diagnostic accuracy was determined for low, intermediate and high thresholds as previously established (ELISA: optical density 0.4, 1.3, and 2.0 respectively; PaGIA: positive/negative, titre of 4, titre of 32; AcuStar HIT-IgG: 1.0 U/ml, 2.8, 9.4) and reproducibility was assessed by repeated measurements. Costs of test determination were calculated taking reagents, controls, and working time of technicians according to Swiss health care system into account.

RESULTS

Data on PaGIA results were available for 171 patients (95.5%), ELISA for 144 patients (80.4%), and AcuStar HIT-IgG for 179 patients (100%). Sensitivity was above 95% for all assays at low and intermediate thresholds. Specificity increased with higher thresholds and was above 90% for all assays with intermediate and high thresholds. Specificity of AcuStar HIT-IgG (92.8%; 95% CI 87.7, 96.2) was significantly higher than PaGIA (83.0%; 95% CI 76.3, 88.5) and higher than ELISA (81.8%, 95% CI 74.2, 88.0) at low threshold (p<0.05). Reproducibility was adequate for all assays. Total costs per test were CHF 51.02 for ELISA, 117.70 for AcuStar HIT-IgG, and 83.13 for PaGIA.

CONCLUSIONS

We observed favourable diagnostic accuracy measures and a high reproducibility for PaGIA and AcuStar HIT-IgG. Implementation into 24-hours-service might improve patient care but the results must be confirmed in other settings and larger populations as well.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Haematology, Oncology, Infectious Diseases, Laboratory Medicine and Hospital Pharmacy (DOLS) > Institute of Clinical Chemistry
04 Faculty of Medicine > Department of Haematology, Oncology, Infectious Diseases, Laboratory Medicine and Hospital Pharmacy (DOLS) > Clinic of Haematology and Central Haematological Laboratory
04 Faculty of Medicine > Pre-clinic Human Medicine > BioMedical Research (DBMR) > Unit Childrens Hospital > Forschungsgruppe Hämatologie (Erwachsene)

UniBE Contributor:

Horn, Michael Peter, Nagler, Michael

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1932-6203

Publisher:

Public Library of Science

Language:

English

Submitter:

Katrin Kölliker-Schütz

Date Deposited:

08 Dec 2017 14:56

Last Modified:

07 Apr 2024 04:03

Publisher DOI:

10.1371/journal.pone.0178289

PubMed ID:

28594835

BORIS DOI:

10.7892/boris.106215

URI:

https://boris.unibe.ch/id/eprint/106215

Actions (login required)

Edit item Edit item
Provide Feedback