Große Hokamp, N; Obmann, Verena Carola; Kessner, R; Laukamp, K R; Persigehl, T; Haneder, S; Ramaiya, N (2018). Improved visualization of hypodense liver lesions in virtual monoenergetic images from spectral detector CT: Proof of concept in a 3D-printed phantom and evaluation in 74 patients. European journal of radiology, 109, pp. 114-123. Elsevier 10.1016/j.ejrad.2018.11.001

[img] Text
1-s2.0-S0720048X18303966-main.pdf - Published Version
Restricted to registered users only
Available under License Publisher holds Copyright.

Download (4MB) | Request a copy

OBJECTIVES

The well-known boost of iodine associated-attenuation in low-keV virtual monoenergetic images (VMI_low) is frequently used to improve visualization of lesions and structures taking up contrast media. This study aimed to evaluate this concept in reverse. Hence to investigate if increased attenuation within the liver allows for improved visualization of little or not-enhancing lesions.

METHODS

A 3D-printed phantom mimicking the shape of a human liver exhibiting a lesion in its center was designed and printed. Both, parenchyma- and lesion-mimic were filled with different solutions exhibiting 80/100/120HU and 0/15/40/60HU, respectively. Further, a total of 74 contrast-enhanced studies performed on a spectral detector CT scanner (SDCT) were included in this retrospective study. Patients had MRI or follow-up proven cysts and/or hypodense metastases. VMI of 40-200 keV as well as conventional images (CI) were reconstructed. ROI were placed in lesion and parenchyma(-mimics) on CI and transferred to VMI. Signal- and contrast-to-noise ratio were calculated (S-/CNR). Further, two radiologists independently evaluated image quality. Data was statistically assessed using ANOVA or Wilcoxon-test.

RESULTS

In phantoms, S/CNR was significantly higher in VMI_low. The cyst-mimic in highly attenuating parenchyma-mimic on CI yielded a CNR of 6.4 ± 0.8; using VMI_40 keV, mildly hypodense lesion-mimic in poorly attenuating parenchyma-mimic exhibited a similar CNR (5.8 ± 0.9; p ≤ 0.05). The same tendency was observed in patients (cyst in CI/metastasis in VMI_40 keV: 4.4 ± 1.2/3.9 ± 1.8; p ≤ 0.05). Qualitative analysis indicated a benefit of VMI_40 keV (p ≤ 0.05).

CONCLUSIONS

VMI_low from SDCT allow for an improved visualization of hypodense focal liver lesions exploiting the concept of contrast blooming in reverse.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Institute of Diagnostic, Interventional and Paediatric Radiology

UniBE Contributor:

Obmann, Verena Carola

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1872-7727

Publisher:

Elsevier

Language:

English

Submitter:

Isabelle Ursula Saurer

Date Deposited:

27 Mar 2019 15:53

Last Modified:

05 Dec 2022 15:25

Publisher DOI:

10.1016/j.ejrad.2018.11.001

PubMed ID:

30527292

Uncontrolled Keywords:

3D-printing Dual energy computed tomography Liver lesions Spectral detector computed tomography Virtual monoenergetic images

BORIS DOI:

10.7892/boris.125597

URI:

https://boris.unibe.ch/id/eprint/125597

Actions (login required)

Edit item Edit item
Provide Feedback