Evaluation of the analytical performance of a portable ion-selective electrode meter for measuring whole-blood, plasma, milk, abomasal-fluid, and urine sodium concentrations in cattle.

Megahed, A A; Hiew, M; Grünberg, W; Trefz, Florian Markus; Constable, P D (2019). Evaluation of the analytical performance of a portable ion-selective electrode meter for measuring whole-blood, plasma, milk, abomasal-fluid, and urine sodium concentrations in cattle. Journal of dairy science, 102(8), pp. 7435-7444. American Dairy Science Association 10.3168/jds.2018-16198

[img] Text
1-s2.0-S0022030219305272-main.pdf - Published Version
Restricted to registered users only
Available under License Publisher holds Copyright.

Download (2MB) | Request a copy

A portable ion-selective electrode (ISE) meter (LAQUAtwin B-722; Horiba Instruments Inc., Irvine, CA) is available for measuring the sodium ion concentration ([Na]) in biological fluids. The objective of this study was to characterize the analytical performance of the ISE meter in measuring [Na] in whole-blood, plasma, milk, abomasal fluid, and urine samples from cattle. Method comparison studies were performed using whole-blood and plasma samples from 106 sick calves and 11 sick cows admitted to a veterinary teaching hospital, 80 milk and 206 urine samples from 16 lactating Holstein-Friesian cows with experimentally induced free water, electrolyte, and acid-base imbalances, and 67 abomasal fluid samples from 7 healthy male Holstein-Friesian calves fed fresh milk with or without an oral electrolyte solution. Deming regression and Bland-Altman plots were used to determine the accuracy of the meter against reference methods. The meter used in direct mode on undiluted samples measured whole-blood [Na] 9.7 mmol/L (7.3%) lower than a direct ISE reference method and plasma [Na] 16.7 mmol/L (12.7%) lower than an indirect ISE reference method. The meter run in direct mode measured milk [Na] 3.1 mmol/L lower and abomasal fluid [Na] 9.0% lower than indirect ISE reference methods. The meter run in indirect mode on diluted samples accurately measured urine [Na] compared with an indirect ISE reference method. We conclude that, after adjustment for the bias determined from Bland-Altman plots, the LAQUAtwin ISE meter provides a clinically useful and low-cost cow-side instrument for measuring [Na] in whole blood, plasma, milk, and abomasal fluid.

Item Type:

Journal Article (Original Article)

Division/Institute:

05 Veterinary Medicine > Department of Clinical Veterinary Medicine (DKV)
05 Veterinary Medicine > Department of Clinical Veterinary Medicine (DKV) > Clinic for Ruminants

UniBE Contributor:

Trefz, Florian Markus

Subjects:

600 Technology > 610 Medicine & health
600 Technology > 630 Agriculture

ISSN:

0022-0302

Publisher:

American Dairy Science Association

Language:

English

Submitter:

Nathalie Viviane Zollinger

Date Deposited:

17 Oct 2019 14:42

Last Modified:

05 Dec 2022 15:31

Publisher DOI:

10.3168/jds.2018-16198

PubMed ID:

31202658

Uncontrolled Keywords:

cattle hypernatremia hyponatremia point-of-care

BORIS DOI:

10.7892/boris.133968

URI:

https://boris.unibe.ch/id/eprint/133968

Actions (login required)

Edit item Edit item
Provide Feedback