Tchétché, Didier; Windecker, Stephan; Kasel, A Markus; Schaefer, Ulrich; Worthley, Stephen; Linke, Axel; Abdel-Wahab, Mohamed; Le Breton, Herve; Søndergaard, Lars; Spence, Mark S; Petronio, Sonia; Baumgartner, Helmut; Hovorka, Tomas; Blanke, Philipp; Reichenspurner, Hermann (2019). 1-Year Outcomes of the CENTERA-EU Trial Assessing a Novel Self-Expanding Transcatheter Heart Valve. JACC. Cardiovascular Interventions, 12(7), pp. 673-680. Elsevier 10.1016/j.jcin.2019.01.231
|
Text
98_30947942 1-Year outcome of the CENTERA-EU.pdf - Published Version Available under License Creative Commons: Attribution-Noncommercial-No Derivative Works (CC-BY-NC-ND). Download (648kB) | Preview |
OBJECTIVES
The purpose of this study is to report the 1-year results of the CENTERA-EU trial.
BACKGROUND
The CENTERA transcatheter heart valve (THV) (Edwards Lifesciences, Irvine, California) is a low-profile (14-F eSheath compatible), self-expanding nitinol valve, with a motorized delivery system allowing for repositionability. The 30-day results of the CENTERA-EU trial demonstrated the short-term safety and effectiveness of the valve.
METHODS
Implantations were completed in 23 centers in Europe, Australia, and New Zealand. Transfemoral access was used in all patients. Echocardiographic outcomes were adjudicated by a core laboratory at baseline, discharge, 30 days, 6 months, and 1 year. Major adverse clinical events were adjudicated by an independent clinical events committee.
RESULTS
Between March 2015 and July 2016, 203 high-risk patients (age 82.7 ± 5.5 years, 67.5% women, 68.0% New York Heart Association functional class III or IV, Society of Thoracic Surgeons score 6.1 ± 4.2%) with severe, symptomatic aortic stenosis underwent transcatheter aortic valve replacement with the CENTERA THV. The primary endpoint of the study was 30-day mortality (1.0%). At 1 year, overall mortality was 9.1%, cardiovascular mortality was 4.6%, disabling stroke was 4.1%, new permanent pacemakers were implanted in 6.5% of patients at risk, and cardiac-related rehospitalization was 6.8%. Hemodynamic parameters were stable at 1 year, with a mean aortic valve gradient of 8.1 ± 4.7 mm Hg, a mean effective orifice area of 1.7 ± 0.42 cm2, and no incidences of severe or moderate aortic regurgitation.
CONCLUSIONS
The CENTERA-EU trial demonstrated mid-term safety and effectiveness of the CENTERA THV, with low mortality, sustained improvements in hemodynamic performances, and low incidence of permanent pacemaker implantations in high-risk patients with symptomatic aortic stenosis. (Safety and Performance of the Edwards CENTERA-EU Self-Expanding Transcatheter Heart Valve [CENTERA-2]; NCT02458560).
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > Department of Cardiovascular Disorders (DHGE) > Clinic of Cardiology |
UniBE Contributor: |
Windecker, Stephan |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
1876-7605 |
Publisher: |
Elsevier |
Language: |
English |
Submitter: |
Nadia Biscozzo |
Date Deposited: |
12 Feb 2020 09:20 |
Last Modified: |
05 Dec 2022 15:35 |
Publisher DOI: |
10.1016/j.jcin.2019.01.231 |
PubMed ID: |
30947942 |
Uncontrolled Keywords: |
CENTERA THV aortic stenosis high-risk patients self-expanding valve transcatheter aortic valve implantation |
BORIS DOI: |
10.7892/boris.139147 |
URI: |
https://boris.unibe.ch/id/eprint/139147 |