Heath, Emily; Mueller, Silvan; Guyer, Gian; Duetschler, Alisha; Eliçin, Olgun; Aebersold, Daniel; Fix, Michael K.; Manser, Peter (2021). Implementation and experimental validation of a robust hybrid direct aperture optimization approach for mixed-beam radiotherapy. Medical physics, 48(11), pp. 7299-7312. Wiley 10.1002/mp.15258
|
Text
mp.15258.pdf - Published Version Available under License Creative Commons: Attribution (CC-BY). Download (10MB) | Preview |
PURPOSE
The objectives of the work presented in this paper were to (1) implement a robust-optimization method for deliverable mixed-beam radiotherapy (MBRT) plans within a previously developed MBRT planning framework; (2) perform an experimental validation of the delivery of robust-optimized MBRT plans; and (3) compare PTV-based and robust-optimized MBRT plans in terms of target dose robustness and organs at risk (OAR) sparing for clinical head and neck and brain patient cases.
METHODS
A robust-optimization method, which accounts for translational setup errors, was implemented within a previously developed treatment planning framework for MBRT. The framework uses a hybrid direct aperture optimization method combining column generation and simulated annealing. A robust plan was developed and then delivered to an anthropomorphic head phantom using the Developer Mode of a TrueBeam linac. Planar dose distributions were measured and compared to the planned dose. Robust-optimized and PTV-based plans were developed for three clinical patient cases consisting of two head and neck cases and one brain case. The plans were compared in terms of the robustness to 5 mm shifts of the target volume dose as well as in terms of OAR sparing.
RESULTS
Using a gamma criterion of 3%/2 mm and a dose threshold of 10%, the agreement between film measurements and dose calculations was better than 97.7% for the total plan and better than 95.5% for the electron component of the plan. For the two head and neck patient cases, the average clinical target volume (CTV) dose homogeneity index (V95%-V107%) over all the considered setup error scenarios was on average 19% lower for the PTV-based plans and it had a larger standard deviation. The robust-optimized plans achieved, on average, a 20% reduction in the OAR doses compared to the PTV-based plans. For the brain patient case, the CTV dose homogeneity index was similar for the two plans, while the OAR doses were 22% lower, on average, for the robust-optimized plan. No clear trend in terms of electron contributions was found across the three patient cases, although robust-optimized plans tended toward higher electron beam energies.
CONCLUSIONS
A framework for robust optimization of deliverable MBRT plans has been developed and validated. PTV-based MBRT were found to not be robust to setup errors, while the dose delivered by the robust-optimized plans were clinically acceptable for all considered error scenarios and had better OAR sparing. This study shows that the robust optimization is a promising alternative to conventional PTV margins for MBRT.