Hecker, Andreas; Eberlein, Sophie C.; Klenke, Frank M. (2022). 3D printed fracture reduction guides planned and printed at the point of care show high accuracy - a porcine feasibility study. Journal of experimental orthopaedics, 9(1), p. 99. Springer 10.1186/s40634-022-00535-2
|
Text
s40634-022-00535-2.pdf - Published Version Available under License Creative Commons: Attribution (CC-BY). Download (1MB) | Preview |
PURPOSE
After surgical treatment of comminuted diaphyseal femoral and tibial fractures, relevant malalignment, especially rotational errors occur in up to 40-50%. This either results in a poor clinical outcome or requires revision surgery. This study aims to evaluate the accuracy of reduction if surgery is supported by 3D guides planned and printed at the point of care.
METHODS
Ten porcine legs underwent computed tomography (CT) and 3D models of femur and tibia were built. Reduction guides were virtually constructed and fitted to the proximal and distal metaphysis. The guides were 3D printed using medically approved resin. Femoral and tibial comminuted diaphyseal fractures were simulated and subsequently reduced using the 3D guides. Postoperative 3D bone models were reconstructed to compare the accuracy to the preoperative planning.
RESULTS
Femoral reduction showed a mean deviation ± SD from the plan of 1.0 mm ± 0.9 mm for length, 0.9° ± 0.7° for varus/valgus, 1.2° ± 0.9° for procurvatum/recurvatum and 2.0° ± 1.7° for rotation. Analysis of the tibial reduction revealed a mean deviation ± SD of 2.4 mm ± 1.6 mm for length, 1.0° ± 0.6° for varus/valgus, 1.3° ± 1.4° for procurvatum/recurvatum and 2.9° ± 2.2° for rotation.
CONCLUSIONS
This study shows high accuracy of reduction with 3D guides planned and printed at the point of care. Applied to a clinical setting, this technique has the potential to avoid malreduction and consecutive revision surgery in comminuted diaphyseal fractures.
LEVEL OF EVIDENCE
Basic Science.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > Department of Orthopaedic, Plastic and Hand Surgery (DOPH) > Clinic of Orthopaedic Surgery ?? 3206CF1930A7491AE053980C5C820121 ?? |
UniBE Contributor: |
Hecker, Andreas, Eberlein, Sophie Charlotte, Klenke, Frank M. |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
2197-1153 |
Publisher: |
Springer |
Language: |
English |
Submitter: |
Pubmed Import |
Date Deposited: |
29 Sep 2022 10:28 |
Last Modified: |
05 Dec 2022 16:25 |
Publisher DOI: |
10.1186/s40634-022-00535-2 |
PubMed ID: |
36166163 |
Uncontrolled Keywords: |
3D print 3D reconstruction Comminuted fracture Diaphyseal fracture Malalignment Malrotation |
BORIS DOI: |
10.48350/173349 |
URI: |
https://boris.unibe.ch/id/eprint/173349 |