Mesbahi, A; Fix, M; Allahverdi, M; Grein, E; Garaati, H (2005). Monte Carlo calculation of Varian 2300C/D Linac photon beam characteristics: a comparison between MCNP4C, GEANT3 and measurements. Applied radiation and isotopes, 62(3), pp. 469-77. Oxford: Pergamon 10.1016/j.apradiso.2004.07.008
Full text not available from this repository.Different codes are used for Monte Carlo (MC) calculations in radiation therapy. In this research, MCNP4C and GEANT3 codes have been compared in calculations of dosimetric characteristics of Varian Clinac 2300C/D. The parameters of influence in the differences seen in dosimetric features were discussed. This study emphasizes that both MCNP4C and GEANT3 MC can be used in radiation therapy computations and their differences in photon spectra calculations have a negligible effect on percentage depth dose computations in radiation therapy.