Impact of a 4th generation iterative reconstruction technique on image quality in low-dose computed tomography of the chest in immunocompromised patients

Laqmani, A.; Buhk, J. H.; Henes, F. O.; Klink, Thorsten; Sehner, S.; von Schultzendorff, H. C.; Hammerle, D.; Nagel, H. D.; Adam, G.; Regier, M. (2013). Impact of a 4th generation iterative reconstruction technique on image quality in low-dose computed tomography of the chest in immunocompromised patients. RöFo. Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 185(8), pp. 749-757. Thieme 10.1055/s-0033-1335577

Full text not available from this repository.

PURPOSE

To determine the image quality of an iterative reconstruction (IR) technique in low-dose MDCT (LDCT) of the chest of immunocompromised patients in an intraindividual comparison to filtered back projection (FBP) and to evaluate the dose reduction capability.

MATERIALS AND METHODS

30 chest LDCT scans were performed in immunocompromised patients (Brilliance iCT; 20-40 mAs; mean CTDIvol: 1.7 mGy). The raw data were reconstructed using FBP and the IR technique (iDose4™, Philips, Best, The Netherlands) set to seven iteration levels. 30 routine-dose MDCT (RDCT) reconstructed with FBP served as controls (mean exposure: 116 mAs; mean CDTIvol: 7.6 mGy). Three blinded radiologists scored subjective image quality and lesion conspicuity. Quantitative parameters including CT attenuation and objective image noise (OIN) were determined.

RESULTS

In LDCT high iDose4™ levels lead to a significant decrease in OIN (FBP vs. iDose7: subscapular muscle 139.4 vs. 40.6 HU). The high iDose4™ levels provided significant improvements in image quality and artifact and noise reduction compared to LDCT FBP images. The conspicuity of subtle lesions was limited in LDCT FBP images. It significantly improved with high iDose4™ levels (> iDose4). LDCT with iDose4™ level 6 was determined to be of equivalent image quality as RDCT with FBP.

CONCLUSION

iDose4™ substantially improves image quality and lesion conspicuity and reduces noise in low-dose chest CT. Compared to RDCT, high iDose4™ levels provide equivalent image quality in LDCT, hence suggesting a potential dose reduction of almost 80%.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Institute of Diagnostic, Interventional and Paediatric Radiology

UniBE Contributor:

Klink, Thorsten

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1438-9029

Publisher:

Thieme

Language:

English

Submitter:

Aisha Stefania Mzinga

Date Deposited:

13 May 2014 09:06

Last Modified:

05 Dec 2022 14:34

Publisher DOI:

10.1055/s-0033-1335577

PubMed ID:

23749649

URI:

https://boris.unibe.ch/id/eprint/52326

Actions (login required)

Edit item Edit item
Provide Feedback