Consecutive series of 226 journey bicruciate substituting total knee replacements: early complication and revision rates

Christen, Bernhard; Neukamp, Michal Sarah; Aghayev, Emin (2014). Consecutive series of 226 journey bicruciate substituting total knee replacements: early complication and revision rates. BMC musculoskeletal disorders, 15, p. 395. BioMed Central 10.1186/1471-2474-15-395

[img]
Preview
Text
Christen_et_al_BMC_Musc_Dis_2014.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (339kB) | Preview

BACKGROUND

The Journey bicruciate substituting (BCS) total knee replacement (TKR) is intended to improve knee kinematics by more closely approximating the surfaces of a normal knee. The purpose of this analysis was to address the safety of Journey BCS knees by studying early complication and revision rates in a consecutive case series.

METHODS

Between December 2006 and May 2011, a single surgeon implanted 226 Journey BCS total knee prostheses in 191 patients (124 women, 67 men) who were eligible for study. Mean age at surgery was 68 years (41-85 years).Outcome measures were early complications and minor and major revision rates. All complications were considered, irrespective of whether conservative treatment or revision was required.

RESULTS

The average implantation time was 3.5 years (range 1.3-5.8 years). Thirty-three complications (14.6% of 226 knees) required minor or major revision surgery in 25 patients. The remaining eight patients were treated conservatively. Sixteen minor revisions were performed in 12 patients. Thirteen major revisions were required in 13 patients, which results in a rate of 1.65 major revisions per 100 component years. The linear trend of the early complication rate by treatment year was not significant (p = .22).Multivariate logistic regression showed no significant predictors for the occurrence of a complication or for revision surgery. A tendency towards higher complication rates was observed in female patients, although it was not significant (p = .066).

CONCLUSIONS

The complication and revision rates of the Journey BCS knee implant are high in comparison with those reported for other established total knee systems. Caution is advised when using this implant, particularly for less experienced knee surgeons.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Pre-clinic Human Medicine > Institute for Evaluative Research into Orthopaedic Surgery

UniBE Contributor:

Neukamp, Michal Sarah, Aghayev, Emin

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1471-2474

Publisher:

BioMed Central

Language:

English

Submitter:

Ives Gerber

Date Deposited:

02 Apr 2015 11:32

Last Modified:

05 Dec 2022 14:44

Publisher DOI:

10.1186/1471-2474-15-395

PubMed ID:

25424844