Accuracy of cartilage-specific 3-Tesla 3D-DESS magnetic resonance imaging in the diagnosis of chondral lesions: comparison with knee arthroscopy

Kohl, Sandro; Meier, Simon; Ahmad, Sufian; Bonel, Harald; Exadaktylos, Aristomenis; Krismer, Anna; Evangelopoulos, Dimitrios Stergios (2015). Accuracy of cartilage-specific 3-Tesla 3D-DESS magnetic resonance imaging in the diagnosis of chondral lesions: comparison with knee arthroscopy. Journal of orthopaedic surgery and research, 10(1), p. 191. BioMed Central 10.1186/s13018-015-0326-1

[img]
Preview
Text
art%3A10.1186%2Fs13018-015-0326-1.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (1MB) | Preview

BACKGROUND

Arthroscopy is considered as "the gold standard" for the diagnosis of traumatic intraarticular knee lesions. However, recent developments in magnetic resonance imaging (MRI) now offer good opportunities for the indirect assessment of the integrity and structural changes of the knee articular cartilage. The study was to investigate whether cartilage-specific sequences on a 3-Tesla MRI provide accurate assessment for the detection of cartilage defects.

METHODS

A 3-Tesla (3-T) MRI combined with three-dimensional double-echo steady-state (3D-DESS) cartilage specific sequences was performed on 210 patients with knee pain prior to knee arthroscopy. Sensitivity, specificity, and positive and negative predictive values of magnetic resonance imaging were calculated and correlated to the arthroscopic findings of cartilaginous lesions. Lesions were classified using the modified Outerbridge classification.

RESULTS

For the 210 patients (1260 cartilage surfaces: patella, trochlea, medial femoral condyle, medial tibia, lateral femoral condyle, lateral tibia) evaluated, the sensitivities, specificities, positive predictive values, and negative predictive values of 3-T MRI were 83.3, 99.8, 84.4, and 99.8 %, respectively, for the detection of grade IV lesions; 74.1, 99.6, 85.2, and 99.3 %, respectively, for grade III lesions; 67.9, 99.2, 76.6, and 98.2 %, respectively, for grade II lesions; and 8.8, 99.5, 80, and 92 %, respectively, for grade I lesions.

CONCLUSIONS

For grade III and IV lesions, 3-T MRI combined with 3D-DESS cartilage-specific sequences represents an accurate diagnostic tool. For grade II lesions, the technique demonstrates moderate sensitivity, while for grade I lesions, the sensitivity is limited to provide reliable diagnosis compared to knee arthroscopy.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Intensive Care, Emergency Medicine and Anaesthesiology (DINA) > University Emergency Center
04 Faculty of Medicine > Department of Orthopaedic, Plastic and Hand Surgery (DOPH) > Clinic of Orthopaedic Surgery

UniBE Contributor:

Kohl, Sandro, Ahmad, Sufian, Exadaktylos, Aristomenis, Krismer, Anna

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1749-799X

Publisher:

BioMed Central

Language:

English

Submitter:

Fabian Röthlisberger

Date Deposited:

23 Mar 2016 10:11

Last Modified:

05 Dec 2022 14:51

Publisher DOI:

10.1186/s13018-015-0326-1

PubMed ID:

26714464

BORIS DOI:

10.7892/boris.75687

URI:

https://boris.unibe.ch/id/eprint/75687

Actions (login required)

Edit item Edit item
Provide Feedback