DESyne novolimus-eluting coronary stent is superior to Endeavor zotarolimus-eluting coronary stent at five-year follow-up: final results of the multicentre EXCELLA II randomised controlled trial

Iqbal, Javaid; Verheye, Stefan; Abizaid, Alexandre; Ormiston, John; de Vries, Ton; Morrison, Lynn; Toyloy, Sara; Fitzgerald, Peter; Windecker, Stephan; Serruys, Patrick W (2015). DESyne novolimus-eluting coronary stent is superior to Endeavor zotarolimus-eluting coronary stent at five-year follow-up: final results of the multicentre EXCELLA II randomised controlled trial. EuroIntervention, 11(6), e1336-e1342. Europa Digital & Publishing 10.4244/EIJY15M10_04

Full text not available from this repository.

AIMS

Newer-generation drug-eluting stents (DES) have been shown to be superior to first-generation DES. Current-generation DES have zotarolimus, everolimus or biolimus as antiproliferative drugs. Novolimus, a metabolite of sirolimus, has been specifically developed to provide efficacy similar to currently available agents at a lower dose and thus requires a lower polymer load. We report the final five-year outcomes of the EXCELLA II trial comparing a zotarolimus-eluting stent (ZES) with a novolimus-eluting stent (NES).

METHODS AND RESULTS

EXCELLA II is a prospective, multicentre, single-blind, non-inferiority clinical trial. Patients (n=210) with a maximum of two de novo lesions in two different epicardial vessels were randomised (2:1) to treatment with either NES (n=139) or ZES (n=71). At five-year follow-up, patients in the NES group had a significantly lower incidence of the patient-oriented (HR 0.53, 95% CI: 0.32-0.87, p=0.013) and device-oriented (HR 0.38, 95% CI: 0.17-0.83, p=0.011) composite endpoints. There was no difference in cardiac death and definite/probable stent thrombosis between the two groups; however, there was a trend towards reduction in myocardial infarction and repeat revascularisation in the NES group at five-year follow-up.

CONCLUSIONS

At five-year follow-up, the incidence of device- and patient-oriented events was significantly lower in the NES group. Further studies, adequately powered for clinical outcomes, are warranted.

TRIAL REGISTRATION

ClinicalTrials.gov number NCT00792753.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Cardiovascular Disorders (DHGE) > Clinic of Cardiology

UniBE Contributor:

Windecker, Stephan

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1774-024X

Publisher:

Europa Digital & Publishing

Language:

English

Submitter:

Judith Liniger

Date Deposited:

07 Apr 2016 08:55

Last Modified:

05 Dec 2022 14:51

Publisher DOI:

10.4244/EIJY15M10_04

PubMed ID:

26465374

URI:

https://boris.unibe.ch/id/eprint/75905

Actions (login required)