Maurer, Martin; Hamm, B; Huppertz, A; Lembcke, A (2015). Ultra-low-dose dual-source CT coronary angiography with high pitch: diagnostic yield of a volumetric planning scan and effects on dose reduction and imaging strategy. British journal of radiology, 88(1049), p. 20140602. British Institute of Radiology 10.1259/bjr.20140602
|
Text
bjr.pdf - Published Version Available under License Publisher holds Copyright. Download (1MB) | Preview |
OBJECTIVE
To evaluate the role of an ultra-low-dose dual-source CT coronary angiography (CTCA) scan with high pitch for delimiting the range of the subsequent standard CTCA scan.
METHODS
30 patients with an indication for CTCA were prospectively examined using a two-scan dual-source CTCA protocol (2.0 × 64.0 × 0.6 mm; pitch, 3.4; rotation time of 280 ms; 100 kV): Scan 1 was acquired with one-fifth of the tube current suggested by the automatic exposure control software [CareDose 4D™ (Siemens Healthcare, Erlangen, Germany) using 100 kV and 370 mAs as a reference] with the scan length from the tracheal bifurcation to the diaphragmatic border. Scan 2 was acquired with standard tube current extending with reduced scan length based on Scan 1. Nine central coronary artery segments were analysed qualitatively on both scans.
RESULTS
Scan 2 (105.1 ± 10.1 mm) was significantly shorter than Scan 1 (127.0 ± 8.7 mm). Image quality scores were significantly better for Scan 2. However, in 5 of 6 (83%) patients with stenotic coronary artery disease, a stenosis was already detected in Scan 1 and in 13 of 24 (54%) patients with non-stenotic coronary arteries, a stenosis was already excluded by Scan 1. Using Scan 2 as reference, the positive- and negative-predictive value of Scan 1 was 83% (5 of 6 patients) and 100% (13 of 13 patients), respectively.
CONCLUSION
An ultra-low-dose CTCA planning scan enables a reliable scan length reduction of the following standard CTCA scan and allows for correct diagnosis in a substantial proportion of patients.
ADVANCES IN KNOWLEDGE
Further dose reductions are possible owing to a change in the individual patient's imaging strategy as a prior ultra-low-dose CTCA scan may already rule out the presence of a stenosis or may lead to a direct transferal to an invasive catheter procedure.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Institute of Diagnostic, Interventional and Paediatric Radiology |
UniBE Contributor: |
Maurer, Martin |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
0007-1285 |
Publisher: |
British Institute of Radiology |
Language: |
English |
Submitter: |
Aisha Stefania Mzinga |
Date Deposited: |
17 Mar 2016 15:04 |
Last Modified: |
05 Dec 2022 14:52 |
Publisher DOI: |
10.1259/bjr.20140602 |
PubMed ID: |
25710210 |
BORIS DOI: |
10.7892/boris.77209 |
URI: |
https://boris.unibe.ch/id/eprint/77209 |