Rausch, Ivo; Füchsel, Frank G; Kuderer, Corinna; Hentschel, Michael; Beyer, Thomas (2016). Radiation exposure levels of routine SPECT/CT imaging protocols. European journal of radiology, 85(9), pp. 1627-1636. Thieme 10.1016/j.ejrad.2016.06.022
Text
Ivo Rausch et al-s2.0-S0720048X1630208X-main.pdf - Published Version Restricted to registered users only Available under License Publisher holds Copyright. Download (1MB) |
AIMS
To assess incremental and total patient exposure from clinical SPECT/CT imaging by means of effective dose estimations with regards to different protocols and SPECT/CT systems.
MATERIALS AND METHODS
Consecutive patient exposure levels were documented prospectively from SPECT/CT operations at three European imaging centers. Documentation included the volume-weighted computed tomography (CT) dose index, the CT dose length product (DLP) and the amount of applied radiopharmaceutical. SPECT/CT examinations were categorized regionally into head, neck, myocardium, thorax, abdomen, extremities and whole-body. Effective dose from the CT (EDCT) was estimated from the DLP using gender specific conversion factors. EDSPECT was estimated from the injected activity levels and corresponding conversion factors (ICRP 106).
RESULTS
This study included 678 SPECT/CT examinations. EDCT per indication and EDSPECT per tracer ranged from 0.01mSv to 7.4mSv and from 1.1mSv to 12.2mSv, respectively. In general, EDSPECT contributed most to overall patient exposure. Total EDSPECT/CT averaged across all protocol categories was 6.7mSv.
CONCLUSION
Total patient exposure from clinical SPECT/CT is 7mSv on average. Individual dose levels vary with the clinical indication and on-site protocol parameters.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Clinic of Nuclear Medicine |
UniBE Contributor: |
Hentschel, Michael |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
0720-048X |
Publisher: |
Thieme |
Language: |
English |
Submitter: |
Franziska Nicoletti |
Date Deposited: |
08 Feb 2017 08:02 |
Last Modified: |
05 Dec 2022 15:01 |
Publisher DOI: |
10.1016/j.ejrad.2016.06.022 |
PubMed ID: |
27501899 |
Uncontrolled Keywords: |
Molecular imaging; Multi-center study; Radiation dosage; SPECT/CT |
BORIS DOI: |
10.7892/boris.93048 |
URI: |
https://boris.unibe.ch/id/eprint/93048 |