Riester, Elena; Majchrzak, Mario; Mühlbacher, Annelies; Tinguely, Caroline; Findeisen, Peter; Kolja Hegel, Johannes; Laimighofer, Michael; Rank, Christopher M.; Schönfeld, Kathrin; Langen, Florina; Niederhauser, Christoph (2021). Multicentre Performance Evaluation of the Elecsys Anti-SARS-CoV-2 Immunoassay as an Aid in Determining Previous Exposure to SARS-CoV-2. Infectious disease and therapy, 10(4), pp. 2381-2397. Springer 10.1007/s40121-021-00504-9
|
Text
Riester2021_Article_MulticentrePerformanceEvaluati.pdf - Published Version Available under License Creative Commons: Attribution-Noncommercial (CC-BY-NC). Download (520kB) | Preview |
Introduction
We performed a multicentre evaluation of the Elecsys® Anti-SARS-CoV-2 immunoassay (Roche Diagnostics), an assay utilising a recombinant protein representing the nucleocapsid (N) antigen, for the in vitro qualitative detection of antibodies to severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2).
Methods
Specificity was evaluated using serum/plasma samples from blood donors and routine diagnostic specimens collected before September 2019 (i.e., presumed negative for SARS-CoV-2-specific antibodies); sensitivity was evaluated using samples from patients with polymerase chain reaction (PCR)-confirmed SARS-CoV-2 infection. Point estimates and 95% confidence intervals (CIs) were calculated. Method comparison was performed versus commercially available assays.
Results
Overall specificity for the Elecsys Anti-SARS-CoV-2 immunoassay (n = 9575) was 99.85% (95% CI 99.75–99.92): blood donors (n = 6714; 99.82%), routine diagnostic specimens (n = 2861; 99.93%), pregnant women (n = 2256; 99.91%), paediatric samples (n = 205; 100.00%). The Elecsys Anti-SARS-CoV-2 immunoassay demonstrated significantly higher specificity versus LIAISON SARS-CoV-2 S1/S2 IgG (99.71% vs. 98.48%), EUROIMMUN Anti-SARS-CoV-2 IgG (100.00% vs. 94.87%), ADVIA Centaur SARS-CoV-2 Total (100.00% vs. 87.32%) and iFlash SARS-CoV-2 IgM (100.00% vs. 99.58%) assays, and comparable specificity to ARCHITECT SARS-CoV-2 IgG (99.75% vs. 99.65%) and iFlash SARS-CoV-2 IgG (100.00% vs. 100.00%) assays. Overall sensitivity for Elecsys Anti-SARS-CoV-2 immunoassay samples drawn at least 14 days post-PCR confirmation (n = 219) was 93.61% (95% CI 89.51–96.46). No statistically significant differences in sensitivity were observed between the Elecsys Anti-SARS-CoV-2 immunoassay versus EUROIMMUN Anti-SARS-CoV-2 IgG (90.32% vs. 95.16%) and ARCHITECT SARS-CoV-2 IgG (84.81% vs. 87.34%) assays. The Elecsys Anti-SARS-CoV-2 immunoassay showed significantly lower sensitivity versus ADVIA Centaur SARS-CoV-2 Total (85.19% vs. 95.06%) and iFlash SARS-CoV-2 IgG (86.25% vs. 93.75%) assays, but significantly higher sensitivity versus the iFlash SARS-CoV-2 IgM assay (86.25% vs. 33.75%).
Conclusion
The Elecsys Anti-SARS-CoV-2 immunoassay demonstrated very high specificity and high sensitivity in samples collected at least 14 days post-PCR confirmation of SARS-CoV-2 infection, supporting its use to aid in determination of previous exposure to SARS-CoV-2.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > Service Sector > Institute for Infectious Diseases > Research 04 Faculty of Medicine > Service Sector > Institute for Infectious Diseases > Infection Serology 04 Faculty of Medicine > Service Sector > Institute for Infectious Diseases |
UniBE Contributor: |
Niederhauser-Lüthi, Christoph Peter |
Subjects: |
600 Technology > 610 Medicine & health 500 Science > 570 Life sciences; biology |
ISSN: |
2193-6382 |
Publisher: |
Springer |
Language: |
English |
Submitter: |
Christoph Peter Niederhauser-Lüthi |
Date Deposited: |
31 Aug 2021 16:27 |
Last Modified: |
12 Dec 2022 18:37 |
Publisher DOI: |
10.1007/s40121-021-00504-9 |
PubMed ID: |
34368915 |
BORIS DOI: |
10.48350/157981 |
URI: |
https://boris.unibe.ch/id/eprint/157981 |