Abou-Ayash, Samir; Schimmel, Martin; Özcan, Mutlu; Oczelik, Burak; Brägger, Urs; Yilmaz, Burak (2021). Trueness and marginal fit of implant-supported complete-arch fixed prosthesis frameworks made of high-performance polymers and titanium: An explorative in-vitro study. Journal of dentistry, 113, p. 103784. Elsevier Science 10.1016/j.jdent.2021.103784

[img]
Preview
Text
1-s2.0-S0300571221002074-main.pdf - Published Version
Available under License Creative Commons: Attribution-Noncommercial-No Derivative Works (CC-BY-NC-ND).

Download (5MB) | Preview

PURPOSE

To investigate the trueness and marginal fit of computer-aided design-computer-aided manufactured (CAD-CAM) complete-arch implant-supported screw-retained fixed prosthesis (CAISFP) made of polyetheretherketone (PEEK), polyetherketoneketone (PEKK) and titanium (Ti) MATERIAL AND METHODS: A typodont model with four implants, their multiunit abutments (MUAs), and MUA scanbodies were digitized by using a laboratory scanner. The generated CAD was used to mill CAISFP frameworks in Ti, PEEK, or PEKK (each n = 10). The frameworks were digitized with an industrial light scanner to superimpose resulting standard tessellation language (STL) file with the CAD file. Deviations at five points at the abutment-framework interface of each of the four abutment sites (1:left first molar, 2:left canine, 3:right canine, 4:right first molar sites) were calculated (trueness). Marginal gaps were measured using the triple scan technique. A nonparametric repeated measures ANOVA by Brunner and Puri with factors being abutment location and material was performed to assess the mean deviations for trueness and mean marginal gaps, followed by Mann-Whitney or exact Wilcoxon Signed-Rank tests (alpha=.05).

RESULTS

Material type significantly affected the trueness (p<0.0001). PEEK had the lowest deviations (0.039 +/-0.01mm) followed by PEKK (0,049 +/-0.009mm), and Ti (0.074 +/-0.011mm). For marginal gaps, only abutment location's effect was significant (p = 0.003). Within PEKK, gaps at abutment 4 were significantly larger, compared with abutments 2 (p = 0.04) and 3 (p = 0.02).

CONCLUSIONS

The trueness of PEEK, PEKK, and Ti frameworks was different after milled. PEEK had the highest trueness. However, the marginal fit of the frameworks was similar and smaller than 90 µm in average.

CLINICAL RELEVANCE

PEEK, PEKK, and Ti complete-arch frameworks had clinically acceptable gaps and may therefore be recommended when their fit is considered. Higher trueness after milling did not result in better marginal fit.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > School of Dental Medicine > Department of Reconstructive Dentistry and Gerodontology

UniBE Contributor:

Abou-Ayash, Samir, Schimmel, Martin, Brägger, Urs, Yilmaz, Burak

ISSN:

0300-5712

Publisher:

Elsevier Science

Language:

English

Submitter:

Lena Augé

Date Deposited:

29 Sep 2021 07:59

Last Modified:

05 Dec 2022 15:53

Publisher DOI:

10.1016/j.jdent.2021.103784

PubMed ID:

34419479

Uncontrolled Keywords:

CAD-CAM Fit analysis High-performance polymers Marginal fit PEEK PEKK Trueness

BORIS DOI:

10.48350/159257

URI:

https://boris.unibe.ch/id/eprint/159257

Actions (login required)

Edit item Edit item
Provide Feedback