Trueness of crowns fabricated by using additively and subtractively manufactured resin-based CAD-CAM materials.

Çakmak, Gülce; Rusa, Ana Maria; Donmez, Mustafa Borga; Akay, Canan; Kahveci, Çiğdem; Schimmel, Martin; Yilmaz, Burak (2024). Trueness of crowns fabricated by using additively and subtractively manufactured resin-based CAD-CAM materials. The journal of prosthetic dentistry, 131(5), pp. 951-958. Elsevier 10.1016/j.prosdent.2022.10.012

[img]
Preview
Text
1-s2.0-S0022391322006904-main.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (2MB) | Preview

STATEMENT OF PROBLEM

Advancements in digital dental technologies have enabled the use of different resin-based materials that can be fabricated either additively or subtractively. However, knowledge on the fabrication trueness of these materials is scarce.

PURPOSE

The purpose of this in vitro study was to investigate the trueness of crowns fabricated by using different resin-based computer-aided design and computer-aided manufacturing (CAD-CAM) materials.

MATERIAL AND METHODS

A complete crown for a mandibular right first molar with a 30-μm cement space was designed in standard tessellation language (STL) format. This master STL (MC-STL) was used to fabricate 40 complete crowns with 4 different resin-based CAD-CAM materials and either additive (Crowntec [MS]) or subtractive techniques (Brilliant Crios [BC], breCAM.monoCOM [PMMA], and G-CAM [GR]; n=10). All crowns were digitized with an intraoral scanner (CEREC Primescan SW 5.2) to generate their STL files (TC-STLs). MC-STL and TC-STLs were transferred into a 3-dimensional analysis software program (Medit Link v2.4.4), and a trueness (overall, external, occlusal, intaglio occlusal, and marginal) analysis was performed by using the root mean square (RMS) method. The Kruskal-Wallis and Dunn tests were performed to analyze data (α=.05).

RESULTS

The test groups had significantly different deviations on all surfaces (P≤.001). MS crowns had higher overall (P≤.007) and external surface (P≤.001) deviations than GR and PMMA crowns, while the differences between GR and PMMA crowns were not significant (P≥.441). BC crowns had higher external surface deviations than GR crowns (P=.005), higher occlusal deviations than GR and MS crowns (P≤.007), and higher intaglio occlusal deviations than GR and MS crowns (P≤.009). However, BC crowns had lower marginal deviations than MS and GR crowns (P≤.018).

CONCLUSIONS

The brand of resin-based CAD-CAM materials affected the trueness of crowns. Additively manufactured crowns (MS) mostly had lower overall and external surface trueness than the other groups. Nevertheless, the deviation values of occlusal, intaglio occlusal, and marginal trueness were generally small; thus, the effect of the tested materials on clinical crown fit may be negligible.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > School of Dental Medicine > Department of Reconstructive Dentistry and Gerodontology
04 Faculty of Medicine > School of Dental Medicine

UniBE Contributor:

Cakmak, Gülce, Dönmez, Mustafa-Borga, Schimmel, Martin, Yilmaz, Burak

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1097-6841

Publisher:

Elsevier

Language:

English

Submitter:

Pubmed Import

Date Deposited:

07 Dec 2022 12:54

Last Modified:

13 May 2024 00:11

Publisher DOI:

10.1016/j.prosdent.2022.10.012

PubMed ID:

36470758

BORIS DOI:

10.48350/175529

URI:

https://boris.unibe.ch/id/eprint/175529

Actions (login required)

Edit item Edit item
Provide Feedback