Impact of intraoral scanner, scanning strategy, and scanned arch on the scan accuracy of edentulous arches: An in vitro study.

Jamjoom, Faris Z; Aldghim, Adhwaa; Aldibasi, Omar; Yilmaz, Burak (2024). Impact of intraoral scanner, scanning strategy, and scanned arch on the scan accuracy of edentulous arches: An in vitro study. The journal of prosthetic dentistry, 131(6), pp. 1218-1225. Elsevier 10.1016/j.prosdent.2023.01.027

[img]
Preview
Text
1-s2.0-S0022391323000690-main.pdf - Published Version
Available under License Creative Commons: Attribution-Noncommercial-No Derivative Works (CC-BY-NC-ND).

Download (924kB) | Preview

STATEMENT OF PROBLEM

The scanning strategy used when making complete arch digital scans affects the accuracy of the scan, and the accuracy of the strategy may be influenced by the scanner used. However, these effects have not been investigated thoroughly with complete arch edentulous scanning.

PURPOSE

The purpose of this in vitro study was to determine the effect of scanning strategies and the scanned arch on the accuracy of complete arch edentulous scans using 2 intraoral scanner (IOS) systems.

MATERIAL AND METHODS

Two IOSs were used (TRIOS 4 and Emerald S) to scan maxillary and mandibular typodonts using 6 scanning strategies (test scans), and conventional impressions of both arches were also made. By using a metrology software program, test scans were superimposed onto a reference scan, and the root mean square (RMS) of the absolute deviation values was calculated to express trueness. The sample with the best trueness was used as reference onto which the remaining samples from the same group were superimposed, and the RMS of the absolute deviation values was calculated to express precision. Statistical modeling was applied using the fixed effects models (α=.05).

RESULTS

The main effects of scanner and strategy significantly impacted the trueness RMS values (P<.001), with significant interactions between them (P=.012). The main effects of scanner, strategy, and arch significantly impacted the precision of RMS values (P=.004), (P=.033), and (P=.023). Conventional impressions and the TRIOS 4 scanner had comparable accuracy, while the Emerald S scanner was inferior to both. P-O-B had the highest overall accuracy and strategy ZZ had the worst. Better precision was found with the maxillary arch.

CONCLUSIONS

The scanner type and scanning strategy significantly impacted the accuracy of the digital scans of completely edentulous arches, with a significant interaction between scanner and strategy. The arch being scanned had a significant effect on scan precision but not on scan trueness.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > School of Dental Medicine > Department of Preventive, Restorative and Pediatric Dentistry
04 Faculty of Medicine > School of Dental Medicine > Department of Reconstructive Dentistry and Gerodontology

UniBE Contributor:

Yilmaz, Burak

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1097-6841

Publisher:

Elsevier

Language:

English

Submitter:

Pubmed Import

Date Deposited:

27 Feb 2023 14:51

Last Modified:

11 Jun 2024 00:11

Publisher DOI:

10.1016/j.prosdent.2023.01.027

PubMed ID:

36841708

BORIS DOI:

10.48350/179269

URI:

https://boris.unibe.ch/id/eprint/179269

Actions (login required)

Edit item Edit item
Provide Feedback