Sedline® Miscalculation of Depth of Anaesthesia Variables in Two Pigs Due to Electrocardiographic Signal Contamination.

Mirra, Alessandro; Hight, Darren; Kovacevic, Alan; Levionnois, Olivier Louis (2023). Sedline® Miscalculation of Depth of Anaesthesia Variables in Two Pigs Due to Electrocardiographic Signal Contamination. Animals, 13(17) MDPI 10.3390/ani13172699

[img]
Preview
Text
animals-13-02699-v2.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (5MB) | Preview

Two young (11-week-old) pigs underwent sole propofol anaesthesia as part of an experimental study. The depth of anaesthesia was evaluated both clinically and using the electroencephalography(EEG)-based monitor Sedline; in particular, the patient state index, suppression ratio, raw EEG traces, and its spectrogram were assessed. Physiological parameters and electrocardiographic activity were continuously monitored. In one pig (Case 1), during the administration of high doses of propofol, the Sedline-generated variables suddenly indicated an increased EEG activity while this was not confirmed by observation of either the raw EEG or its spectrogram. In the second pig (Case 2), a similar event was recorded during euthanasia with systemic pentobarbital. Both events happened while the EEG activity was isoelectric except for signal interferences and synchronous in rhythm and shape with the electrocardiographic activity. The suggestion of increased brain activity based on the interpretation of the Sedline variables was suspected wrong; most probably due to electrocardiographic interferences. In pigs, the patient state index and suppression ratio, as calculated by the Sedline monitor, could be influenced by the electrocardiographic activity contaminating the EEG trace, especially during otherwise isoelectric periods (strong EEG depression). Visual interpretation of the raw EEG and of the spectrogram remains necessary to identify such artefacts.

Item Type:

Journal Article (Further Contribution)

Division/Institute:

04 Faculty of Medicine > Department of Intensive Care, Emergency Medicine and Anaesthesiology (DINA) > Clinic and Policlinic for Anaesthesiology and Pain Therapy > Partial clinic Insel
04 Faculty of Medicine > Department of Intensive Care, Emergency Medicine and Anaesthesiology (DINA) > Clinic and Policlinic for Anaesthesiology and Pain Therapy
05 Veterinary Medicine > Department of Clinical Veterinary Medicine (DKV)
05 Veterinary Medicine > Department of Clinical Veterinary Medicine (DKV) > Small Animal Clinic
05 Veterinary Medicine > Department of Clinical Veterinary Medicine (DKV) > DKV - Anaesthesiology

UniBE Contributor:

Mirra, Alessandro, Hight, Darren Fletcher, Kovacevic, Alan, Levionnois, Olivier

Subjects:

600 Technology > 630 Agriculture

ISSN:

2076-2615

Publisher:

MDPI

Language:

English

Submitter:

Pubmed Import

Date Deposited:

11 Sep 2023 09:21

Last Modified:

29 Jan 2024 09:58

Publisher DOI:

10.3390/ani13172699

PubMed ID:

37684963

Uncontrolled Keywords:

Sedline depth of anaesthesia electrocardiogram electroencephalogram interference patient state index pig signal contamination suppression ratio

BORIS DOI:

10.48350/186198

URI:

https://boris.unibe.ch/id/eprint/186198

Actions (login required)

Edit item Edit item
Provide Feedback