Tidal volume single breath washout of two tracer gases--a practical and promising lung function test

Singer, Florian; Stern, Georgette; Thamrin, Cindy; Fuchs, Oliver; Riedel, Thomas; Gustafsson, Per; Frey, Urs; Latzin, Philipp (2011). Tidal volume single breath washout of two tracer gases--a practical and promising lung function test. PLoS ONE, 6(3), e17588. Lawrence, Kans.: Public Library of Science 10.1371/journal.pone.0017588

[img]
Preview
Text
fetchObject.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (578kB) | Preview

Background

Small airway disease frequently occurs in chronic lung diseases and may cause ventilation inhomogeneity (VI), which can be assessed by washout tests of inert tracer gas. Using two tracer gases with unequal molar mass (MM) and diffusivity increases specificity for VI in different lung zones. Currently washout tests are underutilised due to the time and effort required for measurements. The aim of this study was to develop and validate a simple technique for a new tidal single breath washout test (SBW) of sulfur hexafluoride (SF6) and helium (He) using an ultrasonic flowmeter (USFM).
Methods

The tracer gas mixture contained 5% SF6 and 26.3% He, had similar total MM as air, and was applied for a single tidal breath in 13 healthy adults. The USFM measured MM, which was then plotted against expired volume. USFM and mass spectrometer signals were compared in six subjects performing three SBW. Repeatability and reproducibility of SBW, i.e., area under the MM curve (AUC), were determined in seven subjects performing three SBW 24 hours apart.
Results

USFM reliably measured MM during all SBW tests (n = 60). MM from USFM reflected SF6 and He washout patterns measured by mass spectrometer. USFM signals were highly associated with mass spectrometer signals, e.g., for MM, linear regression r-squared was 0.98. Intra-subject coefficient of variation of AUC was 6.8%, and coefficient of repeatability was 11.8%.
Conclusion

The USFM accurately measured relative changes in SF6 and He washout. SBW tests were repeatable and reproducible in healthy adults. We have developed a fast, reliable, and straightforward USFM based SBW method, which provides valid information on SF6 and He washout patterns during tidal breathing.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Gynaecology, Paediatrics and Endocrinology (DFKE) > Clinic of Paediatric Medicine

UniBE Contributor:

Singer, Florian, Thamrin, Cindy, Fuchs, Oliver, Riedel, Thomas, Frey, Urs Peter, Latzin, Philipp

ISSN:

1932-6203

Publisher:

Public Library of Science

Language:

English

Submitter:

Anette van Dorland

Date Deposited:

04 Oct 2013 14:24

Last Modified:

05 Dec 2022 14:07

Publisher DOI:

10.1371/journal.pone.0017588

PubMed ID:

21423739

Web of Science ID:

000288219100018

BORIS DOI:

10.7892/boris.8387

URI:

https://boris.unibe.ch/id/eprint/8387 (FactScience: 213917)

Actions (login required)

Edit item Edit item
Provide Feedback