Influence of implant angulation and cyclic dislodging on the retentive force of two different overdenture attachments - an in vitro study.

Srinivasan, Murali; Schimmel, Martin; Badoud, Isabelle; Ammann, Patrick; Herrmann, François R; Müller, Frauke (2016). Influence of implant angulation and cyclic dislodging on the retentive force of two different overdenture attachments - an in vitro study. Clinical oral implants research, 27(5), pp. 604-611. Wiley-Blackwell 10.1111/clr.12643

[img] Text
Influence of implant angulation and cyclic dislodging on the retentive force of two different overdenture attachments - an in vitro study..pdf - Published Version
Restricted to registered users only
Available under License Publisher holds Copyright.

Download (728kB)

OBJECTIVE

This in vitro study evaluated the influence of implant angulations on the retentive behavior of two overdenture attachments during cyclic dislodging.

METHODS

Models simulating a two-implant overdenture situation were fabricated. They were divided into five groups based on their simulated implant angulations (Groups: 1 = 0°; 2 = 20°; 3 = 30°; 4 = 40°; and 5 = 60°; n = 90). Each group was further divided into two subgroups based on its attachment system (control attachment: LOCATOR(®) ; test attachment: SFI-Anchor(®) ) except for group 5 which had no LOCATOR(®) group. All models underwent 10,000 insertion-removal cycles in a wet environment. Mean retentive forces were recorded. ANOVA and linear regression models were used for statistical analyses, and the level of significance was at P < 0.05.

RESULTS

The ANOVA model revealed an effect of dislodging cycles for both attachments (P = 0.0070). The linear regression model with repeated measures revealed a significant effect of angulation within the LOCATOR(®) groups (0° vs. 20°: P < 0.0001; 0° vs. 30°: P < 0.0001; 0° vs. 40°: P < 0.0001), but was insignificant within the SFI-Anchor(®) groups (0° vs. 20°: P = 0.544; 0° vs. 30°: P = 0.134; 0° vs. 40°: P = 0.254; 0° vs. 60°: P = 0.979). It further revealed a significant increase in the retentive force between the LOCATOR(®) and the SFI-Anchor(®) (20°: P = 0.041; 30°: P < 0.0001; 40°: P < 0.0001), although there was no significant difference between the attachments at 0° (P = 0.623).

CONCLUSIONS

This study confirms that the retentive behavior of SFI-Anchor(®) is not influenced by implant axial inclination even at angulations of up to 60°. The SFI-Anchor(®) may therefore be particularly indicated for clinical situations with marked implant axial discrepancies.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > School of Dental Medicine > Department of Reconstructive Dentistry and Gerodontology

UniBE Contributor:

Schimmel, Martin

Subjects:

600 Technology > 610 Medicine & health

ISSN:

0905-7161

Publisher:

Wiley-Blackwell

Language:

English

Submitter:

Eveline Carmen Schuler

Date Deposited:

23 Jan 2017 10:03

Last Modified:

05 Dec 2022 15:00

Publisher DOI:

10.1111/clr.12643

PubMed ID:

26085188

Uncontrolled Keywords:

LOCATOR ®, SFI-Anchor®, artificial saliva, dislodging cycles, implant overdentures, in vitro testing, overdentures, retentive force