Srinivasan, Murali; Schimmel, Martin; Badoud, Isabelle; Ammann, Patrick; Herrmann, François R; Müller, Frauke (2016). Influence of implant angulation and cyclic dislodging on the retentive force of two different overdenture attachments - an in vitro study. Clinical oral implants research, 27(5), pp. 604-611. Wiley-Blackwell 10.1111/clr.12643
Text
Influence of implant angulation and cyclic dislodging on the retentive force of two different overdenture attachments - an in vitro study..pdf - Published Version Restricted to registered users only Available under License Publisher holds Copyright. Download (728kB) |
OBJECTIVE
This in vitro study evaluated the influence of implant angulations on the retentive behavior of two overdenture attachments during cyclic dislodging.
METHODS
Models simulating a two-implant overdenture situation were fabricated. They were divided into five groups based on their simulated implant angulations (Groups: 1 = 0°; 2 = 20°; 3 = 30°; 4 = 40°; and 5 = 60°; n = 90). Each group was further divided into two subgroups based on its attachment system (control attachment: LOCATOR(®) ; test attachment: SFI-Anchor(®) ) except for group 5 which had no LOCATOR(®) group. All models underwent 10,000 insertion-removal cycles in a wet environment. Mean retentive forces were recorded. ANOVA and linear regression models were used for statistical analyses, and the level of significance was at P < 0.05.
RESULTS
The ANOVA model revealed an effect of dislodging cycles for both attachments (P = 0.0070). The linear regression model with repeated measures revealed a significant effect of angulation within the LOCATOR(®) groups (0° vs. 20°: P < 0.0001; 0° vs. 30°: P < 0.0001; 0° vs. 40°: P < 0.0001), but was insignificant within the SFI-Anchor(®) groups (0° vs. 20°: P = 0.544; 0° vs. 30°: P = 0.134; 0° vs. 40°: P = 0.254; 0° vs. 60°: P = 0.979). It further revealed a significant increase in the retentive force between the LOCATOR(®) and the SFI-Anchor(®) (20°: P = 0.041; 30°: P < 0.0001; 40°: P < 0.0001), although there was no significant difference between the attachments at 0° (P = 0.623).
CONCLUSIONS
This study confirms that the retentive behavior of SFI-Anchor(®) is not influenced by implant axial inclination even at angulations of up to 60°. The SFI-Anchor(®) may therefore be particularly indicated for clinical situations with marked implant axial discrepancies.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > School of Dental Medicine > Department of Reconstructive Dentistry and Gerodontology |
UniBE Contributor: |
Schimmel, Martin |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
0905-7161 |
Publisher: |
Wiley-Blackwell |
Language: |
English |
Submitter: |
Eveline Carmen Schuler |
Date Deposited: |
23 Jan 2017 10:03 |
Last Modified: |
05 Dec 2022 15:00 |
Publisher DOI: |
10.1111/clr.12643 |
PubMed ID: |
26085188 |
Uncontrolled Keywords: |
LOCATOR ®, SFI-Anchor®, artificial saliva, dislodging cycles, implant overdentures, in vitro testing, overdentures, retentive force |