Bigler, Marius Reto; Stoller, Michael; Praz, Fabien; Siontis, George C M; Grossenbacher, Raphael; Tschannen, Christine; Seiler, Christian (2021). Functional assessment of myocardial ischaemia by intracoronary ECG. Open Heart, 8(1) B M J Group 10.1136/openhrt-2020-001447
|
Text
e001447.full.pdf - Published Version Available under License Creative Commons: Attribution-Noncommercial (CC-BY-NC). Download (2MB) | Preview |
INTRODUCTION
In patients with chronic coronary syndrome, percutaneous coronary intervention targets haemodynamically significant stenoses, that is, those thought to cause ischaemia. Intracoronary ECG (icECG) detects ischaemia directly where it occurs. Thus, the goal of this study was to test the accuracy of icECG during pharmacological inotropic stress to determine functional coronary lesion severity in comparison to the structural parameter of quantitative angiographic per cent diameter stenosis (%S), as well as to the haemodynamic indices of fractional flow reserve (FFR) and instantaneous wave-free ratio (iFR).
METHOD
The primary study endpoint of this prospective trial was the maximal change in icECG ST-segment shift during pharmacological inotropic stress induced by dobutamine plus atropine obtained within 1 min after reaching maximal heart rate(=220 - age). IcECG was acquired by attaching an alligator clamp to the angioplasty guidewire positioned downstream of the stenosis. For the pressure-derived stenosis severity ratios, coronary perfusion pressure and simultaneous aortic pressure were continuously recorded.
RESULTS
There was a direct linear relation between icECG ST-segment shift and %S: icECG=-0.8+0.03*%S (r2=0.164; p<0.0001). There were inverse linear correlations between FFR and %S: FFR=1.1-6.1*10-3*%S (r2=0.494; p<0.0001), and between iFR and %S: iFR=1.27-8.6*10-3*%S (r2=0.461; p<0.0001). Using a %S-threshold of ≥50% as the reference for structural stenosis relevance, receiver operating characteristics-analysis of absolute icECG ST-segment shift during hyperemia showed an area under the curve (AUC) of 0.678±0.054 (p=0.002; sensitivity=85%, specificity=50% at 0.34 mV). AUC for FFR was 0.854±0.037 (p<0.0001; sensitivity=64%, specificity=96% at 0.78), and for iFR it was 0.816±0.043 (p<0.0001;sensitivity=62%, specificity=96% at 0.83).
CONCLUSIONS
Hyperaemic icECG ST-segment shift detects structurally relevant coronary stenotic lesions with high sensitivity, while they are identified highly specific by FFR and iFR.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > Department of Cardiovascular Disorders (DHGE) > Clinic of Cardiology |
Graduate School: |
Graduate School for Cellular and Biomedical Sciences (GCB) |
UniBE Contributor: |
Bigler, Marius Reto, Praz, Fabien Daniel |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
2053-3624 |
Publisher: |
B M J Group |
Language: |
English |
Submitter: |
Andrea Stettler |
Date Deposited: |
20 Jan 2021 15:45 |
Last Modified: |
05 Dec 2022 15:45 |
Publisher DOI: |
10.1136/openhrt-2020-001447 |
PubMed ID: |
33462106 |
Uncontrolled Keywords: |
cardiac catheterisation coronary angiography coronary stenosis electrocardiography percutaneous coronary intervention |
BORIS DOI: |
10.48350/151420 |
URI: |
https://boris.unibe.ch/id/eprint/151420 |