Comuzzi, Luca; Tumedei, Margherita; Romasco, Tea; Petrini, Morena; Afrashtehfar, Kelvin I; Inchingolo, Francesco; Piattelli, Adriano; Di Pietro, Natalia (2022). Insertion Torque, Removal Torque, and Resonance Frequency Analysis Values of Ultrashort, Short, and Standard Dental Implants: An In Vitro Study on Polyurethane Foam Sheets. Journal of Functional Biomaterials, 14(1) MDPI 10.3390/jfb14010010
|
Text
jfb-14-00010.pdf - Published Version Available under License Creative Commons: Attribution (CC-BY). Download (5MB) | Preview |
Short implants were introduced to reduce morbidity, treatment duration, and complex bone regeneration interventions in atrophic jaws and to improve patient-reported outcomes. This study aimed to determine the insertion torque (IT), removal torque (RT), and resonance frequency analysis (RFA) values of ultrashort (3 mm length), short (7 mm length), and standard implants (10 mm length) inserted in 1-, 2-, 3-, and 4-mm thickness polyurethane sheets with densities of 10, 20, and 30 pounds per cubic foot (PCF). Standard-length implants were the gold standard (control). Overall, short-length implant IT values were higher or similar to the control in most experimental conditions. Those inserted into a 3 mm/30 PCF lamina showed the highest IT values, whereas 5 mm diameter ultrashort-length implants inserted into 2 and 3 mm/20 PCF laminas were higher than other implants. RT values followed the same trend and RFA values were more appreciable in short- and standard-length implants in all the scenarios. However, ultrashort-length implants reached a primary stability comparable to that of standard implants in lower thicknesses. In conclusion, although further studies are needed to corroborate this in vitro model with preclinical and clinical studies, our data shed light on short- and ultrashort-length implants geometries to a potential application in critical atrophy of the posterior jaws.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > School of Dental Medicine > Department of Reconstructive Dentistry and Gerodontology |
UniBE Contributor: |
Afrashtehfar, Kelvin Ian |
ISSN: |
2079-4983 |
Publisher: |
MDPI |
Language: |
English |
Submitter: |
Pubmed Import |
Date Deposited: |
23 Jan 2023 09:25 |
Last Modified: |
23 Jan 2023 23:32 |
Publisher DOI: |
10.3390/jfb14010010 |
PubMed ID: |
36662057 |
Uncontrolled Keywords: |
atrophic jaws implant stability insertion torque removal torque short implants ultrashort implants |
BORIS DOI: |
10.48350/177715 |
URI: |
https://boris.unibe.ch/id/eprint/177715 |