Müller, Philipp; Ender, Andreas; Joda, Tim; Katsoulis, Joannis (2016). Impact of digital intraoral scan strategies on the impression accuracy using the TRIOS Pod scanner. Quintessence international, 47(4), pp. 343-349. Quintessenz Verlags-GmbH 10.3290/j.qi.a35524
Full text not available from this repository.OBJECTIVES
Little information is available on the impact of different scan strategies on the accuracy of full-arch scans with intraoral scanners. The aim of this in-vitro study was to investigate the trueness and precision of full-arch maxillary digital impressions comparing three scan strategies.
METHOD AND MATERIALS
Three scan strategies (A, B, and C) were applied each five times on one single model (A, first buccal surfaces, return from occlusal-palatal; B, first occlusal-palatal, return buccal; C, S-type one-way). The TRIOS Pod scanner (3shape, Copenhagen, Denmark) with a color detector was used for these digital impressions. A cast of a maxillary dentate jaw was fabricated and scanned with an industrial reference scanner. This full-arch data record was digitally superimposed with the test scans (trueness) and within-group comparison was performed for each group (precision). The values within the 90/10 percentiles from the digital superimposition were used for calculation and group comparisons with nonparametric tests (ANOVA, post-hoc Bonferroni).
RESULTS
The trueness (mean ± standard deviation) was 17.9 ± 16.4 μm for scan strategy A, 17.1 ± 13.7 μm for B, and 26.8 ± 14.7 μm for C without statistically significant difference. The precision was lowest for scan strategy A (35.0 ± 51.1 μm) and significantly different to B (7.9 ± 5.6 μm) and C (8.5 ± 6.3 μm).
CONCLUSIONS
Scan strategy B may be recommended as it provides the highest trueness and precision in full-arch scans and therefore minimizes inaccuracies in the final reconstruction.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > School of Dental Medicine > Department of Reconstructive Dentistry and Gerodontology |
UniBE Contributor: |
Müller, Philipp, Joda, Tim, Katsoulis, Joannis |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
0033-6572 |
Publisher: |
Quintessenz Verlags-GmbH |
Language: |
English |
Submitter: |
Eveline Carmen Schuler |
Date Deposited: |
23 Jan 2017 09:18 |
Last Modified: |
05 Dec 2022 15:00 |
Publisher DOI: |
10.3290/j.qi.a35524 |
PubMed ID: |
26824085 |
Uncontrolled Keywords: |
accuracy, didgital Impression, intraoral scan, precision, strategy, trueness |
URI: |
https://boris.unibe.ch/id/eprint/91305 |