Technical note: A collision prediction tool using Blender.

Guyer, Gian; Mueller, Silvan; Wyss, Yanick; Bertholet, Jenny; Schmid, Remo; Stampanoni, Marco F M; Manser, Peter; Fix, Michael K (2023). Technical note: A collision prediction tool using Blender. Journal of applied clinical medical physics, 24(11), e14165. American Association of Physicists in Medicine 10.1002/acm2.14165

[img]
Preview
Text
J_Applied_Clin_Med_Phys_-_2023_-_Guyer_-_Technical_note_A_collision_prediction_tool_using_Blender.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (1MB) | Preview

Non-coplanar radiotherapy treatment techniques on C-arm linear accelerators have the potential to reduce dose to organs-at-risk in comparison with coplanar treatment techniques. Accurately predicting possible collisions between gantry, table and patient during treatment planning is needed to ensure patient safety. We offer a freely available collision prediction tool using Blender, a free and open-source 3D computer graphics software toolset. A geometric model of a C-arm linear accelerator including a library of patient models is created inside Blender. Based on the model, collision predictions can be used both to calculate collision-free zones and to check treatment plans for collisions. The tool is validated for two setups, once with and once without a full body phantom with the same table position. For this, each gantry-table angle combination with a 2° resolution is manually checked for collision interlocks at a TrueBeam system and compared to simulated collision predictions. For the collision check of a treatment plan, the tool outputs the minimal distance between the gantry, table and patient model and a video of the movement of the gantry and table, which is demonstrated for one use case. A graphical user interface allows user-friendly input of the table and patient specification for the collision prediction tool. The validation resulted in a true positive rate of 100%, which is the rate between the number of correctly predicted collision gantry-table combinations and the number of all measured collision gantry-table combinations, and a true negative rate of 89%, which is the ratio between the number of correctly predicted collision-free combinations and the number of all measured collision-free combinations. A collision prediction tool is successfully created and able to produce maps of collision-free zones and to test treatment plans for collisions including visualisation of the gantry and table movement.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Haematology, Oncology, Infectious Diseases, Laboratory Medicine and Hospital Pharmacy (DOLS) > Clinic of Radiation Oncology
04 Faculty of Medicine > Department of Haematology, Oncology, Infectious Diseases, Laboratory Medicine and Hospital Pharmacy (DOLS) > Clinic of Radiation Oncology > Medical Radiation Physics

Graduate School:

Graduate School for Cellular and Biomedical Sciences (GCB)

UniBE Contributor:

Guyer, Gian Mauro Carlo, Müller, Silvan Andreas, Bertholet, Jenny, Manser, Peter, Fix, Michael

Subjects:

500 Science > 530 Physics
600 Technology > 610 Medicine & health

ISSN:

1526-9914

Publisher:

American Association of Physicists in Medicine

Language:

English

Submitter:

Pubmed Import

Date Deposited:

05 Oct 2023 15:41

Last Modified:

16 Nov 2023 00:15

Publisher DOI:

10.1002/acm2.14165

PubMed ID:

37782250

Uncontrolled Keywords:

collision prediction non-coplanar radiotherapy treatment planning

BORIS DOI:

10.48350/186864

URI:

https://boris.unibe.ch/id/eprint/186864

Actions (login required)

Edit item Edit item
Provide Feedback