A survey for the use of torque-limiting devices among dental clinicians in Europe.

Yilmaz, Burak; Knapp, Patrick; Cevik, Pinar; Kahveci, Cigdem; Abou-Ayash, Samir (2024). A survey for the use of torque-limiting devices among dental clinicians in Europe. (In Press). Clinical oral implants research Wiley-Blackwell 10.1111/clr.14328

[img]
Preview
Text
Clinical_Oral_Implants_Res_-_2024_-_Yilmaz_-_A_survey_for_the_use_of_torque_limiting_devices_among_dental_clinicians_in.pdf - Published Version
Available under License Creative Commons: Attribution-Noncommercial-No Derivative Works (CC-BY-NC-ND).

Download (2MB) | Preview

PURPOSE

To assess how well torque-limiting devices (TLDs) are known and used by European dentists, and their adherence to screw tightening protocols and screw loosening occurrence through a survey, including the correlation between the dental specialty-of-interest and the recognition, the tightening protocol used, and between the calibration and the occurrence of screw loosening.

MATERIALS AND METHODS

A 10-question survey was distributed to dentists to collect data on their specialty-of-interest, TLD usage, knowledge on TLDs, calibration, the term "preload," tightening speed, tightening protocols used, and occurrence of screw loosening. Pearson test was used for correlation analysis between the specialty-of-interest and the recognition-based questions, the tightening protocol used, and between the calibration and the frequency of screw loosening.

RESULTS

Of 422 respondents, 24% calibrated their TLDs, 27% knew the term "preload," 76% selected the correct location to read on TLDs, and 6% was aware of the effect of tightening speed. The correlation between the specialty-of-interest and the recognition-based questions was nonsignificant (p < .05) but was significant for used tightening protocol (p < .001). The correlation between the calibration and the occurrence of screw loosening was nonsignificant (p = 0.16). Tightening protocols' effect on screw loosening was similar, which was mostly observed less than once a year (p < .001).

CONCLUSIONS

A lack in dentists' knowledge was found on calibration, the term preload, and the effect of tightening speed, which were not impacted by the dentists' specialty-of-interest, which affected the preferred tightening protocol. The tightening protocol and calibration did not impact the occurrence of screw loosening, which was mostly observed less than once a year.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > School of Dental Medicine > Department of Reconstructive Dentistry and Gerodontology
04 Faculty of Medicine > School of Dental Medicine > Department of Preventive, Restorative and Pediatric Dentistry

UniBE Contributor:

Yilmaz, Burak, Abou-Ayash, Samir

Subjects:

600 Technology > 610 Medicine & health

ISSN:

0905-7161

Publisher:

Wiley-Blackwell

Language:

English

Submitter:

Pubmed Import

Date Deposited:

07 Aug 2024 11:43

Last Modified:

08 Aug 2024 00:01

Publisher DOI:

10.1111/clr.14328

PubMed ID:

39105457

Uncontrolled Keywords:

calibration dental abutments dental prosthesis implant‐supported torque