Yilmaz, Burak; Knapp, Patrick; Cevik, Pinar; Kahveci, Cigdem; Abou-Ayash, Samir (2024). A survey for the use of torque-limiting devices among dental clinicians in Europe. (In Press). Clinical oral implants research Wiley-Blackwell 10.1111/clr.14328
|
Text
Clinical_Oral_Implants_Res_-_2024_-_Yilmaz_-_A_survey_for_the_use_of_torque_limiting_devices_among_dental_clinicians_in.pdf - Published Version Available under License Creative Commons: Attribution-Noncommercial-No Derivative Works (CC-BY-NC-ND). Download (2MB) | Preview |
PURPOSE
To assess how well torque-limiting devices (TLDs) are known and used by European dentists, and their adherence to screw tightening protocols and screw loosening occurrence through a survey, including the correlation between the dental specialty-of-interest and the recognition, the tightening protocol used, and between the calibration and the occurrence of screw loosening.
MATERIALS AND METHODS
A 10-question survey was distributed to dentists to collect data on their specialty-of-interest, TLD usage, knowledge on TLDs, calibration, the term "preload," tightening speed, tightening protocols used, and occurrence of screw loosening. Pearson test was used for correlation analysis between the specialty-of-interest and the recognition-based questions, the tightening protocol used, and between the calibration and the frequency of screw loosening.
RESULTS
Of 422 respondents, 24% calibrated their TLDs, 27% knew the term "preload," 76% selected the correct location to read on TLDs, and 6% was aware of the effect of tightening speed. The correlation between the specialty-of-interest and the recognition-based questions was nonsignificant (p < .05) but was significant for used tightening protocol (p < .001). The correlation between the calibration and the occurrence of screw loosening was nonsignificant (p = 0.16). Tightening protocols' effect on screw loosening was similar, which was mostly observed less than once a year (p < .001).
CONCLUSIONS
A lack in dentists' knowledge was found on calibration, the term preload, and the effect of tightening speed, which were not impacted by the dentists' specialty-of-interest, which affected the preferred tightening protocol. The tightening protocol and calibration did not impact the occurrence of screw loosening, which was mostly observed less than once a year.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > School of Dental Medicine > Department of Reconstructive Dentistry and Gerodontology 04 Faculty of Medicine > School of Dental Medicine > Department of Preventive, Restorative and Pediatric Dentistry |
UniBE Contributor: |
Yilmaz, Burak, Abou-Ayash, Samir |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
0905-7161 |
Publisher: |
Wiley-Blackwell |
Language: |
English |
Submitter: |
Pubmed Import |
Date Deposited: |
07 Aug 2024 11:43 |
Last Modified: |
08 Aug 2024 00:01 |
Publisher DOI: |
10.1111/clr.14328 |
PubMed ID: |
39105457 |
Uncontrolled Keywords: |
calibration dental abutments dental prosthesis implant‐supported torque |
BORIS DOI: |
10.48350/199539 |
URI: |
https://boris.unibe.ch/id/eprint/199539 |