Molinero-Mourelle, Pedro; Abou-Ayash, Samir; Brägger, Urs; Schimmel, Martin; Özcan, Mutlu; Yilmaz, Burak; Buser, Ramona; Al-Haj Husain, Nadin (2024). Load bearing capacity of 3-unit screw-retained implant-supported fixed dental prostheses with a mesial and distal cantilever on a single implant: A comparative in vitro study. Journal of the mechanical behavior of biomedical materials, 151, p. 106395. Elsevier 10.1016/j.jmbbm.2024.106395
|
Text
1-s2.0-S1751616124000274-main.pdf - Published Version Available under License Creative Commons: Attribution (CC-BY). Download (2MB) | Preview |
OBJECTIVES
To assess the mechanical durability of monolithic zirconia implant-supported fixed dental prostheses (iFDP) design on one implant, with a distal and a mesial extension cantilever bonded to a titanium base compared to established designs on two implants.
MATERIALS AND METHODS
Roxolid Tissue level (TL), and tissue level x (TLX) implants were used to manufacture screw-retained 3-unit iFDPs (n = 60, n = 10 per group), with following configurations (X: Cantilever; I: Implant, T: Test group, C: Control group): T1: X-I-X (TL); T2: X-I-X (TLX); T3: I-I-X (TL); T4: I-I-X (TLX); C1: I-X-I (TL); C2: I-X-I (TLX). The iFDPs were thermomechanically aged and subsequently loaded until fracture using a universal testing machine. The failure load at first crack (Finitial) and at catastrophic fracture (Fmax) were measured and statistical evaluation was performed using two-way ANOVA and Tukey's post-hoc tests.
RESULTS
The mean values ranged between 190 ± 73 and 510 ± 459 N for Finitial groups, and between 468 ± 76 and 1579 ± 249 N for Fmax, respectively. Regarding Finitial, neither the implant type, nor the iFDP configuration significantly influenced measured failure loads (all p > 0.05). The choice of implant type did not show any significant effect (p > 0.05), while reconstruction design significantly affected Fmax data (I-I-Xa < X-I-Xb < I-X-Ic) (p < 0.05). The mesial and distal extension groups (X-I-X) showed fractures only at the cantilever extension site, while the distal extension group (I-I-X) showed one abutment and one connector fracture at the implant/reconstruction interface.
CONCLUSION
Results suggest that iFDPs with I-X-I design can be recommended regardless of tested implant type followed by the mesial and distal extension design on one implant abutment (X-I-X).
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > School of Dental Medicine > Department of Reconstructive Dentistry and Gerodontology 04 Faculty of Medicine > School of Dental Medicine > Department of Preventive, Restorative and Pediatric Dentistry |
UniBE Contributor: |
Molinero Mourelle, Pedro, Abou-Ayash, Samir, Brägger, Urs, Schimmel, Martin, Yilmaz, Burak, Buser, Ramona, Al-Haj Husain, Nadin |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
1878-0180 |
Publisher: |
Elsevier |
Language: |
English |
Submitter: |
Pubmed Import |
Date Deposited: |
22 Jan 2024 15:38 |
Last Modified: |
03 Feb 2024 00:16 |
Publisher DOI: |
10.1016/j.jmbbm.2024.106395 |
PubMed ID: |
38244420 |
Uncontrolled Keywords: |
Dental implant Dental restoration failure Fatigue Implant-supported dental prosthesis Load-bearing capacity Prosthodontics Screw-retained |
BORIS DOI: |
10.48350/191956 |
URI: |
https://boris.unibe.ch/id/eprint/191956 |